GC Analysis of Oxygen, Nitrogen, and Carbon Dioxide on Carboxen®-1010 PLOT

GC Analysis of Oxygen, Nitrogen, and Carbon Dioxide on Carboxen®-1010 PLOT

Conditions

column Carboxen-1010 PLOT, 30 m x 0.53 mm I.D. (25467)
oven 35 C (7.0 min), 24 C/min to 225 C
carrier gas helium, 3.0 ml/min
sample Oxygen at 20% and carbon dioxide at 1.0% in nitrogen
injection 10 µL, direct using a sample valve, 150 C
inj. temp. 200 C
detector TCD, 230 C, Methanizer, 375 C, and FID, 230 C in series

Description

Categories Analytical Chromatography, GC Applications, Volatiles, Gases
Featured Industry Chemicals and Industrial Polymers
Legal Information Carboxen is a registered trademark of Sigma-Aldrich Co. LLC
Other Notes Complementary products:
Syringes
Vials
Gas Purification and Gas Management
Inlet Liners, Glass Wool, and Specialized Hand Tools
Septa and Specialized Hand Tools
Column Ferrules, Nuts, and Specialized Hand Tools
suitability application for GC

Materials

     
Related Links