GC Analysis of α-Pinene Enantiomers on Astec® CHIRALDEX® B-DM

GC Analysis of α-Pinene Enantiomers on Astec® CHIRALDEX® B-DM

Conditions

column Astec CHIRALDEX B-DM, 30 m x 0.25 mm I.D., 0.12 µm (77023AST)
oven 50 °C
carrier gas helium, 30 psi
sample 4 mg/mL in methylene chloridepeak 1: l-alpha-pinene (S(-)-alpha-pinene)peak 2: d-alpha-pinene (R(+)-alpha-pinene)
injection 1 µL, 80:1 split
inj. temp. 250 °C
detector FID, 250 °C

Description

Analysis Note Elution order reversal (d,l) on CHIRALDEX G-DP.
Categories Analytical Chromatography, GC Applications, Terpenes, Enantiomers