GC Analysis of Water in Ethanol (60%Standard) on Watercol™ 1460

GC Analysis of Water in Ethanol (60%Standard) on Watercol™ 1460

Conditions

column Watercol 1460, 30 m x 0.25 mm I.D., 0.20 µm (29695-U)
oven 60 °C (12 min)
carrier gas helium, 26 cm/sec
sample water at 80% (v/v) in ethanol
injection 0.1 µL, 100:1 split
liner 4 mm I.D., split type, cup design
inj. temp. 250 °C
detector TCD, 200 °C

Description

Analysis Note The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. This application shows a calibration standard (60% water in ethanol) on Watercol 1460.