Attention:

Certain features of Sigma-Aldrich.com will be down for maintenance the evening of Friday August 18th starting at 8:00 pm CDT until Saturday August 19th at 12:01 pm CDT.   Please note that you still have telephone and email access to our local offices. We apologize for any inconvenience.

US EPA Method 8082: GC Analysis of Aroclor 1016/1260 (PCBs) on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

US EPA Method 8082: GC Analysis of Aroclor 1016/1260 (PCBs) on SLB®-5ms (15 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

Conditions

column SLB-5ms, 15 m x 0.10 mm I.D., 0.10 µm (28466-U)
oven 80 C (0.5 min), 50 C/min to 200 C, 35 C/min to 360 C (2 min)
carrier gas hydrogen, 40 cm/sec constant
sample Aroclor standard mix 1 (46846-U) diluted to 500 ppb / 50 ppb (Aroclors / surrogates) in n-hexane
injection 2 µL, splitless (0.75 min)
liner 4 mm I.D., single taper
inj. temp. 225 C
detector ECD, 360 C

Description

Categories Analytical Chromatography, GC Applications, Dioxins/PCBs/PBDEs
Featured Industry Environmental
Legal Information SLB is a registered trademark of Sigma-Aldrich Co. LLC
Other Notes Complementary products:
Syringes
Vials
Gas Purification and Gas Management
Inlet Liners, Glass Wool, and Specialized Hand Tools
Septa and Specialized Hand Tools
Column Ferrules, Nuts, and Specialized Hand Tools
suitability application for GC

Analyte

  • 2,4,5,6-Tetrachloro-m-xylene
  • Aroclor 1016
  • Aroclor 1260
  • Decachlorobiphenyl

Materials

     
Related Links