HPLC Analysis of Apotransferrin Tryptic Digest on BIOshell™ A160 Peptide C18, Comparison of Different Flow Rates and Pressures

HPLC Analysis of Apotransferrin Tryptic Digest on BIOshell™ A160 Peptide C18, Comparison of Different Flow Rates and Pressures

Conditions

column BIOSHELL A160 Peptide C18, 15 cm x 4.6 mm I.D., 5 µm particles (67015-U)
column temp. 60 °C
mobile phase [A] water/0.1% TFA [B]: acetonitrile/0.1% TFA
gradient 5-50%B in 30 minutes; 5-50%B in 15 minutes;
flow rate 2 mL/min or 4/mL minute
pressure 126 bar, 250 bar
sample 1 mg/mL in water/0.1% TFA
injection 15 µL
detector UV, 215 nm

Description

Analysis Note BIOshell columns packed with five micron Fused-Core particles feature modest back pressures, which allows the use of very high flow rates as is demonstrated in the two chromatograms, one of which was obtained at 2mL/min, the other at 4mL/min. Note that at 4mL/min the gradient time was reduced in half to keep the gradient volume constant. At a flow rate of 4mL/min the back pressure for this 15cm x 4.6mm column is 250 bar (approx. 3500 psi), which illustrates that these high efficiency columns can be operated on standard HPLC systems.