HPLC Analysis of Arotinolol Enantiomers on Astec® CHIROBIOTIC® T

HPLC Analysis of Arotinolol Enantiomers on Astec® CHIROBIOTIC® T

Conditions

column CHIROBIOTIC T, 25 cm x 4.6 mm I.D., 5 µm particles (12024AST)
column temp. 25 C
mobile phase [A] acetonitrile; [B] methanol; [C] acetic acid; [D] triethylamine; (55:45:0.3:0.2, A:B:C:D)
flow rate 2 ml/min
sample 5 mg/mL in methanol
injection 5 µl
detector UV, 230 nm

Description

Categories Analytical Chromatography, Chiral Applications, Antihypertensives, Pharmaceutical (small molecule), All Pharmaceutical Applications
Featured Industry Pharmaceutical (small molecule)
Legal Information CHIROBIOTIC is a registered trademark of Sigma-Aldrich Co. LLC, Astec is a registered trademark of Sigma-Aldrich Co. LLC
suitability application for HPLC

Materials

     
Related Links