HPLC Analysis of Procainamide and N-Acetylprocainamide on Discovery® HS F5

HPLC Analysis of Procainamide and N-Acetylprocainamide on Discovery® HS F5

Conditions

column Discovery HS F5, 15 cm x 4.6 mm I.D., 5 µm particles (567516-U)
column temp. 30 °C
mobile phase [A] 25 mM ammonium phosphate (pH 7): [B] acetonitrile; (65:35, A:B)
flow rate 1 mL/min
sample 50 µg/mL
injection 5 µL
detector UV, 280 nm

Description

Categories Analytical Chromatography, Pharmaceutical Analysis, Antiarrythmic
Featured Industry Pharmaceutical (small molecule)
Legal Information Discovery is a registered trademark of Sigma-Aldrich Co. LLC
suitability application for HPLC

Materials

     