HPLC Analysis of Terfenadine Enantiomers on Astec® Cellulose DMP

HPLC Analysis of Terfenadine Enantiomers on Astec® Cellulose DMP

Conditions

column Astec Cellulose DMP, 15 cm x 4.6 mm I.D., 5 µm particles (51098AST)
column temp. 40 C
mobile phase [A] heptane; [B] ethanol; [C] diethylamine; (98:2:0.1, A:B:C)
flow rate 0.5 ml/min
sample 1 g/L in methanol
injection 3 µl
detector UV, 230 nm