IC Separation of Chloride, Fluoride, Nitrite, Bromide, Nitrate, Phosphate, and Sulfate Ions on Metrohm Metrosep A Supp 5

IC Separation of Chloride, Fluoride, Nitrite, Bromide, Nitrate, Phosphate, and Sulfate Ions on Metrohm Metrosep A Supp 5

Conditions

column Metrosep A Supp 5, 15 cm x 4 mm I.D., 5 µm particles (Metrohm 6.1006.520)
column temp. 25 °C
mobile phase [A] 3.2 mM sodium carbonate; [B] 1.0 mM sodium bicarbonate
flow rate 0.7 mL/min
sample chloride, fluoride (2 mg/L), nitrite (5 mg/L), bromide, nitrate, phosphate, sulfate (10 mg/L)
injection 20 µL
detector conductivity

Description

Categories Analytical Chromatography, Anions, Ion chromatography
Featured Industry Chemicals and Industrial Polymers
Environmental
Food and Beverages
General Analytical
suitability application for HPLC

Materials

     
Related Links