LC/MS Analysis of 3-Methoxytyramine and Metanephrines on SeQuant ZIC-HILIC

LC/MS Analysis of 3-Methoxytyramine and Metanephrines on SeQuant ZIC-HILIC

Conditions

column SeQuant ZIC-HILIC, 10 cm x 2.1 mm, 3.5 µm, 100 Ã… particles (1.50441.0001)
column temp. 25 °C
mobile phase 5 mM ammonium formate with 0.1% formic acid in 85:15, acetonitrile:water
flow rate 0.5 mL/min
pressure 3500 psi (241 bar)
sample 200 ng/mL in acetonitrile
injection 1 µL
detector MS, ESI+, combined SIR

Description

Categories Analytical Chromatography, Catecholamines
Featured Industry Clinical
suitability application for LC-MS

Materials

     
Related Links