LC/MS Analysis of Pesticides on Ascentis® C18 (5 cm x 2.1 mm x 3 μm)

LC/MS Analysis of Pesticides on Ascentis® C18 (5 cm x 2.1 mm x 3 μm)

Conditions

column Ascentis C18, 5 cm x 2.1 mm I.D., 3 µm particles (581300-U)
column temp. 35 °C
mobile phase [A] 0.1% ammonium acetate, pH unadjusted; [B] acetonitrile
gradient 10 to 80% B in 10 min
flow rate 0.2 mL/min
sample 1 µg/mL each in 90:10 water:acetonitrile
injection 5 µL
detector MS, (+) ESI, Selected Ion Recording Mode

Description

Categories Analytical Chromatography, Food & Beverage Analysis, Environmental Analysis, Pesticide and Metabolite Residues, Organics, Pesticides & Herbicides, Chromatograms, Applications (Chromatograms), HPLC Chromatograms
Featured Industry Environmental
Legal Information Ascentis is a registered trademark of Sigma-Aldrich Co. LLC
suitability application for HPLC

Materials

     