Analytical Method: Phosphorus in textile finish solutions

Instruments and Reagents:

Cat.No. 173016  Spectroquant® VIS Spectrophotometer Prove 100 or

Cat.No. 173017  Spectroquant® UV/VIS Spectrophotometer Prove 300 or

Cat.No. 173018  Spectroquant® UV/VIS Spectrophotometer Prove 600 or

Cat.No. 109748  Spectroquant® Photometer NOVA 30 or

Cat.No. 109751  Spectroquant® Photometer NOVA 60 or

Cat.No. 173632  Spectroquant® Colorimeter Move 100

Cat.No. 114946  Rectangular cells 10 mm and/or

Cat.No. 114947  Rectangular cells 20 mm and/or

Cat.No. 114944  Rectangular cells 50 mm
Cat.No. 114543  Spectroquant® Phosphate Cell Test
Cat.No. 114729  Spectroquant® Phosphate Cell Test
Cat.No. 116754  Water for analysis
Thermoreactor
 
Sample preparation:
Cleaning solutions contain, beside from pebble brine (highly disperged silicic acid), condensed phosphates and soluble phosphorus acidic ester. In a standard volumetric flask dilute the sample according to the expected
P-content 1:100 to 1:200 with water for analysis and filter it.
 
Analysis:
Determine with the above mentioned test kits.
 
Calculation:
Multiply the analysis value with the factor of dilution of the sample The result is equal to phosphorus (total) content in the original sample.