HPLC Analysis of Peptides on Discovery® BIO Wide Pore C18 (Gradient #1)

HPLC Analysis of Peptides on Discovery® BIO Wide Pore C18 (Gradient #1)

Conditions

column Discovery BIO Wide Pore C18, 15cm x 4.6mm I.D., 5mum particles (568222-U)
column temp. 30 C
mobile phase [A] (80:20), 0.1% TFA in water:0.1% TFA in acetonitrile; [B] (66:34), 0.1% TFA in water:0.1% TFA in acetonitrile
gradient 0-100% B in 14 min after 1 minute delay
flow rate 1 ml/min
sample Sigma Peptide Mix (Cat. No. P2693)
injection 10 µl
detector UV, 220 nm