48 hours has been validated as the shortest time required for a successful Sterikon incubation. Up to 7 days is validated as can be seen in the Application note: https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/588/304/sterikon-plus-incubation-time-an1800en-ms.pdf
Select a Size
| Size/SKU | Availability | Price |
|---|---|---|
15 amp | Estimated to ship onAugust 05, 2026fromMilliporeSigma Canada Ltd. | $108.00 |
100 amp | Estimated to ship onAugust 05, 2026fromMilliporeSigma Canada Ltd. | $587.00 |
About This Item
Quality Segment
agency
EP, USP
form
liquid
feature
ready-to-use
packaging
ampule of 2 mL
application(s)
microbiology
storage temp.
2-8°C
suitability
Bacillus spp.
General description
Application
Analysis Note
Number of viable spores per ampoule: 5.0E+05 - 1.0E+07
Growth (35 °C): no growth
Growth (60 °C): growth
Performance test (6 min, 121 °C): Survival time 6 min/121 °C
Performance test (15 min, 121 °C): Kill time 15 min/121 °C
D value (121 °C): 1.5 - 2.0 min
Z value: 7.0 - 10.0 °C
Identity: Geobacillus stearo-thermophilus ATCC 7953
The determination of the D-value is based on the Stumbo-Murphy-Cochran method.
The identity control is carried out via MicroSeq-System.
Organism: Geobacillus stearothermophilus ATCC 7953 (sporulation optimized).
Ampoules: 2 ml.
Storage temperature: +2° to +8°C.
Other Notes
Legal Information
1 of 1
This Item | |||
|---|---|---|---|
| feature ready-to-use | feature ready-to-use | feature - | feature hydrophilic |
| packaging ampule of 2 mL | packaging - | packaging pkg of 5 mL (100 mg) | packaging pack of 12 ea |
| form liquid | form liquid (ready-to-use) | form crystalline | form - |
| suitability Bacillus spp. | suitability Bacillus geobacillus | suitability - | suitability - |
| application(s) microbiology | application(s) food and beverages, microbiology | application(s) - | application(s) - |
| agency EP, USP | agency EP, USP | agency - | agency - |
Storage Class
12 - Non Combustible Liquids
wgk
WGK 1
flash_point_f
Not applicable
flash_point_c
Not applicable
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
-
I need to know the recommended tolerance for bioindicator checks related to incubation duration. best regard
1 answer-
Helpful?
-