The 10 kDa membrane suitable for this stirred cell is PLGC07610.
Select a Size
| Size/SKU | Availability | Price |
|---|---|---|
1 ea | Please contact Customer Service for Availability | $3,430.00 |
About This Item
Quality Segment
material
Nylon assembly (cap and tube fitting), Nylon stand (retaining), Tygon® tube (filtrate), anodized aluminum stand (retaining), polyacetal assembly (stirring), polyethylene tube (pressure), polysulfone assembly (stirring), polysulfone holder (cylinder & membrane), silicone O-ring
product line
Amicon®
manufacturer/tradename
Amicon®
parameter
400 mL process volume
technique(s)
protein clean-up: suitable (desalting), protein purification: suitable (concentration)
H
18.7 cm (7.4 in.)
filter diam.
76 mm
filtration area
41.8 cm2
shipped in
ambient
Application
Biochem/physiol Actions
Other Notes
Legal Information
1 of 1
This Item | |||
|---|---|---|---|
| shipped in ambient | shipped in ambient | shipped in ambient | shipped in ambient |
| Quality Level 200 | Quality Level 200 | Quality Level 200 | Quality Level 200 |
| manufacturer/tradename Amicon® | manufacturer/tradename Amicon® | manufacturer/tradename Amicon® | manufacturer/tradename Amicon® |
| filtration area 41.8 cm2 | filtration area 28.7 cm2 | filtration area 13.4 cm2 | filtration area 41.8 cm2 |
| parameter 400 mL process volume | parameter 200 mL process volume | parameter 50 mL process volume | parameter 400 mL process volume |
| product line Amicon® | product line Amicon® | product line Amicon® | product line Amicon® |
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
-
Does this model have a 10 kDa membrane?
1 answer-
Helpful?
-
-
Could you please provide more information about the troubleshooting for a stirred cell that is not holding pressure and has poor concentrate recovery?
1 answer-
On September 7, 2021, the Eyre User Guide provided troubleshooting information on pages 8 and 9. It addressed the symptom of a stirred cell not holding pressure and poor concentrate recovery. The cause identified was a membrane that was not sealed, leading to sample bypass around the membrane. The corrective action recommended was to ensure that the O-ring is free of dirt or debris and undamaged.
Helpful?
-
-
What is the minimum volume that can be concentrated using stirred cells? Additionally, what is the minimum volume that can be recovered from the stirred cell?
1 answer-
The stirred cells do not have a specified minimum concentrate volume, as this is typically dependent on the sample. According to the user guide, the minimum concentrate volume must be determined empirically by the user based on sample composition and stirring speed.
In theory, it is possible to concentrate to dryness; however, this is highly dependent on the sample and membrane used.
Additionally, the specification chart includes a hold-up volume, which indicates the volume of sample that is 'lost' to the device. For detailed hold-up volume information, please refer to page 10 of the user guide.
Helpful?
-