GC Analysis of Gasoline Residue in Arson Samples on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

Materiali
Prodotti correlati
N° Catalogo
Descrizione
Determinazione del prezzo
used together
N° Catalogo
Descrizione
Determinazione del prezzo
Unità fibra SPME polidimetilsilossano (PDMS)
100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, red hubCONDITIONS
SPME fiber
polydimethylsiloxane phase fiber, 100 μm headspace sampling (20 min) 10 sec desorption (splitless mode) (57300-U)
column
100% methyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28046-U)
oven
35 °C (2 min) to 220 °C at 10 °C/min, hold 2 min, to 300 °C at 30 °C/min, hold 5 min
inj. temp.
splitless (closed 3 min), 220 °C (2 mm I.D. injector liner)
detector
FID, 300 °C
carrier gas
helium, 1 mL/min (split 50:1)