Product Name

Oxygen Cell Test, photometric, 0.5-12.0 mg/L (O2), Spectroquant®

product line

Spectroquant®

Quality Level

usage

sufficient for 25 tests

specific analyte(s)

oxygen

measuring range

0.5-12.0 mg/L (O2)

technique(s)

photometry: suitable

application(s)

aquaculture
boiler and cooling water
environmental
ground and surface water
swimming pools

compatibility

for use with Spectroquant® Move 100
for use with Spectroquant® Nova 60 A
for use with Spectroquant® Prove 100 (plus)
for use with Spectroquant® Prove 300 (plus)
for use with Spectroquant® Prove 600 (plus)

detection method

photometric (mod. Winkler method)

storage temp.

15-25°C

Compare Similar Items

View Full Comparison

Show Differences

1 of 4

This Item
1.192511.147311.00594
specific analyte(s)

oxygen

specific analyte(s)

-

specific analyte(s)

hydrogen peroxide

specific analyte(s)

aluminium

detection method

photometric (mod. Winkler method)

detection method

photometric (Iron reduction)

detection method

photometric (Titanyl sulfate)

detection method

photometric ( Chromazurol S )

technique(s)

photometry: suitable

technique(s)

photometry: suitable

technique(s)

photometry: suitable

technique(s)

photometry: suitable

measuring range

0.5-12.0 mg/L (O2)

measuring range

0.020-0.500 mg/L (DEHA), 0.027-0.667 mg/L (Carbohy), 0.05-1.32 mg/L (Hydro), 0.08-1.95 mg/L (ISA), 0.09-2.17 mg/L (MEKO)

measuring range

0.25-5.0 mg/L (H2O2), 2.0-20.0 mg/L (H2O2)

measuring range

0.02-0.50 mg/L (Al)

compatibility

for use with Spectroquant® Move 100, for use with Spectroquant® Prove 100, for use with Spectroquant® Prove 600, for use with Spectroquant® Nova 60 A

compatibility

for use with Spectroquant® Move 100, for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100, for use with Spectroquant® Prove 300, for use with Spectroquant® Prove 600

compatibility

for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100, for use with Spectroquant® Prove 300, for use with Spectroquant® Prove 600

compatibility

for use with Spectroquant® Move 100, for use with Spectroquant® Nova 60 A, for use with Spectroquant® Prove 100, for use with Spectroquant® Prove 300, for use with Spectroquant® Prove 600

product line

Spectroquant®

product line

Spectroquant®

product line

Spectroquant®

product line

Spectroquant®

General description

Method: photometric 0.5 - 12.0 mg/l O₂ Spectroquant®

The method is analogous to DIN EN 25813-21.This Spectroquant® Oxygen Cell Test allows the accurate quantification of the oxygen concentration in aqueous samples. Method applied: Oxygen oxidizes manganese(II) to manganese(III). In acidic solution, manganese(III) reacts with Titriplex® II to form a red complex which is determined photometrically (modified Winkler method).The Spectroquant® Cell Tests come with prefilled 16 mm round cells and all the required reagents to perform the analysis according to the instruction leaflet provided. All Spectroquant® Cell and Reagent Test Kits are equipped with the unique Live ID (2D barcode) which allows seamless method recognition and contains essential information such as lot number, expiry date, and automatic calibration updates.

Legal Information

SPECTROQUANT is a registered trademark of Merck KGaA, Darmstadt, Germany
Titriplex is a registered trademark of Merck KGaA, Darmstadt, Germany

pictograms

CorrosionExclamation mark

signalword

Danger

Storage Class

8B - Non-combustible, corrosive hazardous materials


Certificates of Analysis (COA)

Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.

Already Own This Product?

Find documentation for the products that you have recently purchased in the Document Library.

Visit the Document Library

Questions

  1. hello, I was wondering if I can use any spectrophotometer to measure with this kit. And also I would like to know at which waveband I should measure. Thanks

    1 answer
    1. Yes, other spectrophotometers can work with this kit as well. The recommended wavelength for measurement is 500 nm.