The Amicon Ultra devices are non-sterile. However, many customers rinse with 70% IPA then discard, let dry, then introduce a sample to the product.
Select a Size
| Pack Size | SKU | Availability | Price |
|---|---|---|---|
| 8 pkg | Estimated to ship TODAY | $82.30 | |
| 24 pkg | Estimated to ship TODAY | $200.00 | |
| 96 pkg | Estimated to ship TODAY | $758.00 |
About This Item
Quality Segment
material
polyethylene cap, polyethylene liner, polypropylene tube, regenerated cellulose membrane, regenerated cellulose support, styrene-butadiene housing
sterility
non-sterile
product line
Amicon®
packaging
pkg of 24 pkg, pkg of 8 pkg, pkg of 96 pkg
manufacturer/tradename
Amicon® Ultra
sustainability
Greener Alternative Product
parameter
4 mL sample volume
technique(s)
protein extraction: suitable, protein purification: suitable
L
124 mm (4.88 in.)
diam.
17.3 mm
filtration area
3 cm2
volume
4 mL
working volume
≤4 mL , 4 mL
pore size
3 kDa MWCO
greener alternative category
shipped in
ambient
General description
The vertical design and available membrane surface area provide fast sample processing, high sample recovery (typically greater than 90% of dilute starting solution), and the capability for 80-fold concentration. Solute polarization and subsequent fouling of the membrane are minimized by the vertical design, and a physical deadstop in the filter device prevents spinning to dryness and potential sample loss.
The concentrate is collected from the filter device sample reservoir using a pipetter, while the ultrafiltrate is collected in the provided centrifuge tube. The device can be spun in a swinging bucket or fixed angle rotor. With their vertical membrane design and deadstop, Amicon® Ultra 15 mL filters deliver the highest recovery with short spin times.
Application
- Ultrafiltration with 3,000 Dalton molecular weight cutoff Ultracel® membrane.
- Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids, or microorganisms.
- Purification of macromolecular components found in tissue culture extracts or cell lysates and protein removal prior to HPLC.
- Concentration of dilute or pre-purified proteins from column eluents.
- Desalting, buffer exchange and protein dialysis.
- Size exclusion and protein fractionation.
Features and Benefits
- Ultracel® regenerated cellulose membrane with 3 kDa NMWCO delivers retentate recovery of greater than 90%.
- Vertical membrane reduces concentration polarization for ultra-fast spin times (as fast as 5 minutes).
- Heat-sealed membrane minimizes downstream extractables.
- 100% integrity tested for reliable performance.
- Convenient sample monitoring with translucent housing and volume gradations.
- Direct pipette sample access eliminates processing step to recover concentrate.
- High concentration factors of 80–100X.
Physical form
Legal Information
1 of 1
This Item | |||
|---|---|---|---|
| pore size 3 kDa MWCO | pore size 3 kDa MWCO | pore size 3 kDa MWCO | pore size 3 kDa MWCO |
| parameter 4 mL sample volume | parameter 2 mL sample volume | parameter 15 mL sample volume | parameter 0.5 mL sample volume |
| manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra |
| material polyethylene cap, polypropylene tube, regenerated cellulose support, polyethylene liner, styrene-butadiene housing, regenerated cellulose membrane | material polypropylene tube (for collection), regenerated cellulose membrane, styrene-butadiene housing | material polyethylene cap, polyethylene liner, polypropylene tube, regenerated cellulose membrane, regenerated cellulose support, styrene-butadiene housing | material polypropylene cap, polypropylene tube, regenerated cellulose membrane, styrene-butadiene housing |
| working volume ≤4 mL | working volume ≤2 mL | working volume ≤15 mL | working volume ≤500 μL , 0.5 mL |
| packaging pkg of 24 pkg, pkg of 96 pkg, pkg of 8 pkg | packaging pkg of 24 pkg | packaging pkg of 24 pkg, pkg of 8 pkg, pkg of 96 pkg | packaging pkg of 24 pkg, pkg of 500 pkg, pkg of 8 pkg, pkg of 96 pkg |
Still not finding the right product?
Explore all of our products under Amicon® Ultra Centrifugal Filter, 3 kDa MWCO
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Protocols
El proteoma plasmático humano es un enorme recurso para el descubrimiento y la investigación de biomarcadores.
The human plasma proteome is a vast resource for biomarker discovery and investigation.
Articles
Explore ultrafiltration membrane principles and benefits for protein sample preparation, including solvent exchange and sample concentration.
Filtration methods and membrane filter selection in nanoparticle purification and production.
Global Trade Item Number
| SKU | GTIN |
|---|---|
| UFC800396 | 08436037126024 |
| UFC800308 | 04053252265938 |
| UFC800324 | 08436037126017 |
-
Good day, I would like to know if these concentrators are sterile and the recovered proteins are suitable for downstream cell culture assays. Thank you!
1 answer-
Helpful?
-
-
What could be the reason for no protein recovery (in the concentrate or flow through) while using the UFC800324 to purify protein? Could the protein may be absorbed in the filter?
1 answer-
The low or no protein recovery issue may be attributed to over-concentration of the protein, leading to precipitation. The User Guide provides guidance on maximizing sample recovery: Low sample recovery in the concentrate may be due to adsorptive losses, over-concentration, or passage of sample through the membrane. Adsorptive losses depend upon solute concentration, its hydrophobic nature, temperature and time of contact with filter device surfaces, sample composition, and pH. To minimize losses, remove concentrated samples immediately after centrifugal spin. If the starting sample concentration is high, monitor the centrifugation process in order to avoid over-concentration of the sample. Over-concentration can lead to precipitation and potential sample loss. If the sample appears to be passing through the membrane, choose a lower MWCO Amicon device.
Helpful?
-
-
What is max g force these filters can take? i.e. what sort of rpm or g force should I spin at?
1 answer-
The UFC8003 protocol states to spin at 7,500 × g maximum for approximately 10–40 minutes. See the image below for further details.
Helpful?
-
-
Can we centrifuge these tubes in a normal 15 mL swinging bucket centrifuge?
1 answer-
Yes, the device can be used in centrifuges with a swinging bucket or fixed-angle rotor.