This Molecular Sieve 5A Moisture Trap can reduce moisture in the gas stream to final concentrations of less than 0.1 ppm.
Skip To
Product Name
Molecular Sieve 5A Moisture Trap, volume 750 cc, 1/4 in. fitting
volume
750 cc
fitting
1/4 in. fitting
Looking for similar products? Visit Product Comparison Guide
1 of 4
This Item | 23992 | 503118 | 20618 |
|---|---|---|---|
| volume 750 cc | volume 750 cc | volume - | volume 200 cc |
| fitting 1/4 in. fitting | fitting 1/2 in. fitting | fitting 1/8 in. fitting | fitting 1/4 in. fitting |
Application
General description
Dimensions:
- 200 cc Trap: 26 1/4 in. (67 cm) long × 1 in. (2.5 cm) diameter
- 750 cc Trap: 18 in. (45.7 cm) long × 2 3/8 in. (6 cm) diameter
- S-Trap: 7 1/2 in. (19 cm) long x 5 9/16 in. (14.1 cm) wide x 1/2 in. (13 mm) diameter [total bed length: 19 3/4 in. (50.2 cm); the extended bed length ensures prolonged contact between the gas and the adsorbent and provides greater working capacity]
related product
signalword
Warning
hcodes
pcodes
Hazard Classifications
STOT RE 2 Inhalation
target_organs
Lungs
Storage Class
13 - Non Combustible Solids
wgk
nwg
flash_point_f
Not applicable
flash_point_c
Not applicable
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
-
How dryness of the outlet gas can this moisture trap provide in ppm moisture level?
1 answer-
Helpful?
-
-
Hi, We just purchased the Molecular Sieve 5A. What flow rate could we expect to get with an inlet pressure of 200 psi? If we connected another Sieve 5a in parallel would we double the flow rate? Thanks Vic
1 answer-
The 200 cc Molecular Sieve 5A Trap is rated for a maximum flow rate of 2,000 cc/min at an inlet pressure of 250 psig/17.2 bar. The 750 cc Molecular Sieve 5A Trap is rated for a maximum flow rate of 10,000 cc/min at an inlet pressure of 250 psig/17.2 bar. In theory, if the same inlet pressure can be maintained for two traps run in parallel, the effective flow rate measured from the combined outlets would double. In practice, this flow rate may not be exactly doubled due to some uncontrollable variables between a single trap and multi-trap system.
Helpful?
-
-
What is the Department of Transportation shipping information for this product?
1 answer-
Transportation information can be found in Section 14 of the product's (M)SDS.To access the shipping information for this material, use the link on the product detail page for the product.
Helpful?
-
-
How do I refill Product No. 23991, Molecular Sieve 5A Water Vapor Trap?
1 answer-
Product No. 23991 at one end has a two piece fitting. Remove the 1/8 in NPT fitting and empty contents into an appropriate container for proper disposal. Clean and replace any Teflon sealing material. Refill the trap with four of Product No. 20298.
Helpful?
-