Capillary Column Ferrule Selection and Installation

Robert F. Wallace

Reporter US Volume 27.3

Gas chromatography injector with ferrules and column connection showing sample introduction into the system

Introduction

To ensure a leak-tight seal, an ideal gas chromatography (GC) ferrule should accommodate column outside diameter (O.D.) variations, seal effectively with minimal torque, and resist sticking to the column or fittings. The proper selection of any GC ferrule depends on compatibility with the column material, operating temperature, connection requirements, and variations in column diameter. Our Supeltex® ferrules and CapSeal Bullet® ferrules deliver leak-free across a wide range of applications.

DescriptionMax
temp.
CompositionCharacteristicsApplications
Supeltex® M-1 Ferrules250 °CCeramic-filled PTFE

High reusability. Isothermal use only. Limited use inside oven.

Glass packed columns at low temperatures.
Supeltex® M-1P Ferrules
250 °C100% PTFESoft, completely inert, and have very low friction. Hydrophobic and commonly used in liquid applications. These can be reused, susceptible to air penetrationGlass packed columns at low temperatures.
Supeltex® M-2 Ferrules
350 °CDuPont VESPEL SP-1
(100% polyimide)
Hard polymeric material that tends to seals permanently to capillary columns.
Not reusable and tend to shrink through repeated heating and cooling cycles. 
Isothermal operation. Work great on metal or glass with long lifetimes. May be drilled with a pin vise. Traditional used outside the GC oven.
Supeltex® M-2A Ferrules
400 °CDuPont VESPEL SP-21
(85% polyimide/15% graphite)
Seals at 1/4-turn past finger-tight. High reusability. Should be retightened after initial temperature cycles to avoid leaks. Graphite reduces the shrinkage associated with thermal cycling. Ideal for metal or glass (non-stick). General purpose for capillary or packed columns. Dominate ferrule inside the GC oven.
Supeltex® M-4 Ferrules450 °CFlexible graphiteSeals at 1/4-turn past finger-tight. Maximum sealing surface contact. Do not shrink but are very soft and most be handled with care. Do not stick to the column, should not be used with MSD or ECDs as not air tight. Can be drilled with a pin vise.Suitable for high temperature and cool on-column applications. High-temperature and cool on-column applications.
CapSeal Bullet®450 °CGraphite in an aluminum baseSeals at 1/8-turn past finger-tight. Reusable. A special end taper reduces graphite extrusion into fitting. Suitable for high temperature and cool on-column applications. High-temperature and cool on-column applications.
SilTite® Ferrules500 °CMetal FerrulesStrong permanent airtight seals. These are reusable. High temperature tolerance.Developed specifically for GC-MS systems.
Table 1.Ferrule properties and applications

Ferrule Selection by Detector

The choice of ferrule material can also depend on the detector being used in the gas chromatography system. Some detectors have specific requirements to ensure optimal performance and prevent contamination or air leakage.

Detector type

Recommended ferrule materials

Mass Spectrometric Detector (MSD)

Vespel/Graphite (85/15) or SilTite® metal ferrules. 100% Vespel is also suitable. Avoid 100% graphite as it is porous to air and can cause elevated background.

Electron Capture Detector (ECD)

Vespel/Graphite (85/15) or SilTite® metal ferrules. Avoid 100% graphite ferrules as they allow air permeation which interferes with detector sensitivity.

Flame Ionization Detector (FID)
and other general detectors

Any ferrule material is suitable, but graphite or Vespel/Graphite (85/15) are commonly used for general-purpose applications.

Table 2.Recommended ferrule materials by detector type

Choosing the Correct Ferrule I.D. for Capillary Columns

After selecting a ferrule material appropriate for the capillary column and application, the next step is to choose a ferrule size that matches the column. Before tightening, the ferrule should fit snugly around the column, as loose ferrules may crack or damage the column wall during tightening. This damage can lead to leaks or column breakage after repeated heating and cooling cycles. 

Recommended Diameters

Column

Ferrules

0.10-0.25 mm

0.4 mm I.D.

0.25 mm

0.4 mm I.D.

0.32 mm

0.5 mm I.D.

0.53 mm

0.8 mm I.D.

Table 3.Recommended ferrule I.D. based on GC column diameters

Proper Ferrule Installation

Correct installation of the ferrule is critical to maintaining column performance and ensuring a reliable, leak-free connection. The following steps are recommended

  1. If sealed, cut the capillary column end squarely using a dedicated tool such as a Capillary Cleaving™ Tool (23740-U) or Shortix® Tubing Cutter (21386-U).
  2. Select a ferrule with an internal diameter that matches the column outer diameter.
  3. Keep the column end pointed towards the floor during cutting the end and while installing the nut and ferrule. This allows column fragments and ferrule particles to fall away from the column.
  4. Install the nut and ferrule onto the column according to the instrument manufacturer’s instructions.
  5. After installing the ferrule, remove potential debris by cutting approximately one inch from the column end. This step should be performed each time a ferrule is installed.
  6. Inspect the cut surface using a magnifier(e.g. EISCO-PH0502B, not available everywhere, or HS15092d) to ensure it is square and clean.
  7. Follow the instrument manufacturer’s specified insertion distances when positioning a fused silica  column in the injector or detector port, as improper placement can adversely affect column performance.
  8. Insert the column into the inlet and detector ends, tighten fittings by hand, and use a wrench for final tightening while avoiding excessive force.
  9. Verify all connections using an electronic leak detector (GOW-MAC® Miniature Leak Detector, 110 V, 22807) to ensure the system is leak-free.

Choosing the Correct Ferrule and Nut Combination

Different ferrule designs used in gas chromatography for sealing column connections and ensuring leak-free operation

Representative ferrule designs: short, long, and general (left to right).

Ferrule Designs

Ferrules are available in different designs to ensure compatibility with specific GC fittings and nuts.

Short design ferrules fit:

  • Original nuts supplied with Agilent GCs

Long design ferrules fit:

  • MSD source nuts for Agilent GCs
  • Original nuts supplied with PerkinElmer® GCs
  • Original nuts supplied with Varian GCs

General purpose ferrules fit:

  • Supelco® ferrule nut adapters for Agilent GCs
  • 1/16 inch compression nuts for PerkinElmer® GCs

Column OD, Ferrule ID

Shape

M-2

Vespel

M-2A

Vespel/Graphite 85/15

M-4

Graphite

Capseal Bullet®

Graphite Aluminum base

0.10 -0.25 mm,

0.4 mm ferrule

Short

GF100004A-10

GF851504S-10, 24803-U

24811-U

23864

 

24807-U

 

23867

General

 

224672-hole

  
 

503258

22498

 

 

22474

22480-U

 

Long

GF100004L-10

GF851504L-10, 24826-U

28025-U

 

 

28022-U

  

0.32 mm,

0.5 mm ferrule

Short

GF100005A-10

GF851505S-10

24809-U

23865

 

24806-U

24813-U

23868

General

 

22461

  
  

22462

23481

  

22412

23486

Long

GF100005L-10

GF851505L-10, 24824-U

28026-U

23489

0.53 mm,

0.8 mm ferrule

Short

GF100008A-10

GF851508S-10

24808-U

23866

 

24804-U

24812-U

23869

General

 

22489

20628

23482

  

22479

23487

Long

GF100008L-10

GF851508L-10

28027-U

23495

 

28024-U

  

Blank

Short GF85150XX-10  
General 

22488

  
Table 4.Ferrule selection for capillary GCcolumns by size, design, and material

Ferrule Nuts

Gas chromatography ferrule nuts used for securing columns and fittings in GC systems

(A) 24833-U, (B) 22470-U, (C) 28034-U

Type

Ferrule design

Use

Cat. No.

Agilent nut

Short

Agilent Inj. and non-MS detectors

24833-U

Supelco® ferrule nut adapter

General

Agilent Inj. and non-MS detectors

22470-U

Agilent MSD

Long

Agilent [MS detector] and PerkinElmer

28034-U

Perkin Elmer nut

Long

Agilent [MS detector] and PerkinElmer

28034-U

Table 5.Nut types for capillary GC columns

SilTite® Ferrules and SilTite® FingerTite® Ferrules

We offer two innovative metal ferrule systems for gas chromatography, SilTite® ferrules and SilTite® FingerTite® ferrules. Both the products represent significant improvements over traditional graphite and composite ferrules, providing better sealing and temperature tolerance, but differing in key design features and installation methods. This comparison highlights the essential differences between these two high-performance GC connection systems.

Feature

SilTite® ferrules

SilTite® FingerTite® ferrules

Installation method

Requires wrench for tightening

Finger-tight only (no tools required)

System components

Standard metal ferrules and nuts

Special female nuts, adapters, and
installation tool

Seal type

Strong permanent airtight seal

Leak-free and maintenance-free seal

Temperature limit

Well above GC oven capacity

No temperature limit (same as SilTite® ferrules)

Column protection

Good, but potential for overtighten

Cannot generate enough force with fingers to break columns

Reusability

Reusable with care

Column with ferrule can be removed/reinserted multiple times

Optimized for

General GC applications

Specifically designed for GC-MS applications

Ease of use

Requires technical skill

Simplified installation process, saves time

Material

Metal (stainless steel)

Metal (stainless steel)

Special feature

Same expansion rate as nut to prevent leaks

Adapters enable finger-force connection

Table 6.Overview of SilTite® ferrules and SilTite® FingerTite® ferrules

Column OD, Ferrule ID

SilTite® ferrule

SilTite® FingerTite® ferrule

0.10-0.25 Column,
0.4 mm Ferrule

GF010025MS-10

GF00004H-10

0.32 Column,
 0.5 mm Ferrule

GF00032MS-10

GF00005H-10

0.53 column,
0.7 mm Ferrule

GF04553MS-10

GF00007H-10

Blank Ferrule*

 

GF0000XX-10

1/32' Column,
1/32' Ferrule

GF00080MS-10

 
*Blank Ferrules (also known as no-hole or blind ferrules) may be used to seal unused ports on injectors, detectors, or manifolds
Siltite® metal nuts and ferrules for gas chromatography providing durable and leak-tight column connections

SilTite® metal nuts and ferrules.

Description

Cat. No.

SilTite® Metal Nuts for Agilent Split / Splitless Injectors, PK5

GFMN001INJ-5

SilTite® Metal Nuts for Agilent GC-MS (10-32 thread), PK5

GFMN001MS-5

SilTite® FingerTite® Female Nut ; Pk.10

GF0000FF-10

Siltite® fingertite® starter kit, injector fitting, and FID adapter installation on fused silica tubing for gas chromatography systems

(A) SilTite® FingerTite® Generic Starter Kit on Fused Silica Tubing, (B) SilTite® FingerTite® Injector for Agilent, (C) SilTite® FingerTite® FID Adapter Instillation, (D) SilTite® FingerTite® Instillation Tool

Kit

Cat. No.

Kit components

SilTite® FingerTite® Generic Starter Kit

 

 

 

GFKIT001S

5x FingerTite® nuts

(GF000FF-10)

10x FingerTite® ferrules

(GF00004H-10)

2x MS adapters

Measure Tool

Ferrule install tool (for injector and MS interface)

SilTite® FingerTite® 0.25 mm Starter Kit

 

 

GFTKIT025C

5x Fingertite® nuts

(GF0000FF-10)

10x FingerTite® ferrules

(GF00004H-10)

2x MS adapter

Measure Tool

SilTite® FingerTite® 0.32 mm Starter Kit

 

 

GFKIT032C

5x FingerTite® nuts

(GF000FF-10)

10x FingerTite® ferrules

(GF00005H-10)

2x MS adapter

Measure Tool

SilTite® FingerTite® 0.53 mm Starter Kit
for Agilent Injector

GFKIT053C

10x FingerTite® nuts

(GF000FF-10)

5x FingerTite® ferrules

(GF00007H-10)

2x Agilent Injector Port adapters

2x Agilent Base seal

Measure Tool

Exploded diagram of Siltite® fingertite® injection port assembly showing nut, ferrule, seal, adapter, and base components for GC connection

Figure 1.Exploded diagram of the injection port SilTite® FingerTite® assembly. (1) Original injector nut (not supplied) (2) Adapter-base seal injection port (3) base seal washer (4) nut injection port (5) 0.4 mm SilTite® FingerTite® ferrule (6) SilTite® FingerTite® nut.

Exploded diagram of Siltite® fingertite® FID detector assembly showing adapter, ferrule, nut, and sealing components for GC systems


Figure 2.
Exploded diagram of the FID detector SilTite® FingerTite® assembly. (1) Adapter-FID port (2) 0.4 mm SilTite® FingerTite® ferrule (3) SilTite® FingerTite® nut (4) FID ferrule (GVF) (5) Nut FID

Packed GC and Reducing Ferrules

M-1 ferrules used in gas chromatography for sealing column connections

M-1 ferrule

M-1P ferrules for gas chromatography fittings ensuring secure and leak-tight connections

M-1P ferrule

M-2 ferrules used in gas chromatography systems for column and fitting connections

M-2 ferrule

M-2A ferrules for gas chromatography providing reliable sealing in analytical systems

M-2A ferrule

M-4 ferrules used in gas chromatography for stable and leak-free column connections

M-4 ferrule

Fitting, Ferrule ID

Composition

M-1
PTFE w/ceramic

M-1P
PTFE

M-2
100% Vespel

M-2A
Vespel/Graphite 85/15

M-4

Graphite

0.75 mm, 1.0 mm ferrule    

22494

0.75 mm, 1.2 mm ferrule

   

22459

22460

1/16" to 1/16"

22386

23862-U

GF0100UUL-10

 

22487-U

22495-U

1/8" to 1/8"

  

22476

22472

22491

1/8" to 1/16"

   

22484-U

 

1/4" to 1/4"

22086-U

22087-U

29024-U

22475

22481

22471

22492

22478

1/4" to 1/16"

   22486 
Table 10.Ferrule sizes for packed GC columns by fitting, I.D., and material composition (Pk. of 10 if not indicated otherwise).

Tips and Best Practices

  • For oxygen-sensitive detectors such as MSD or ECD, always use non-porous ferrule materials like Vespel/Graphite or SilTite® metal ferrules.
  • Vespel and Vespel/Graphite ferrules shrink upon heating and may require retightening after initial temperature cycles to prevent leaks.
  • Precondition graphite-containing ferrules before use with sensitive detectors to minimize background signals.
  • Replace ferrules when reinstalling a column rather than reusing existing ones.
  • Metal ferrules provide the highest temperature stability.

For column selection, refer to the GC Capillary Column Selection Guide.

Conclusion

Selecting the appropriate ferrule for your analysis does not need to be complex. Following a straightforward selection approach can help reduce the risk of leaks and prevent issues such as ferrules becoming stuck in fittings. Proper ferrule selection and installation can save significant time and effort each time a new capillary GC column is installed.