Biosimilar Comparability Workflow for Intact Mass Analysis Using Trastuzumab as Model Protein

Geoffrey Rule1, Sundaram Palaniswamy2, Uma Sreenivasan3, Paresh Tank4, Fatima D’Souza4

1Bellefonte, PA, USA, 2Bangalore, India, 3Round Rock, TX, USA, 4ZRAS, Zelle Biotechnology Pvt. Ltd., Thane, India

Introduction

Biosimilar mAb drugs have surged in the last decade due to patent expiration of innovator drugs, offering affordable care to patients. Biosimilars, while sharing amino acid sequences with innovators, can differ structurally due to post-translational modifications (PTMs) resulting from changes in manufacturing processes. These variations, like N-glycosylation and alterations to termini, pose challenges in proving comparability.1

Thorough characterization of therapeutic mAbs is vital for safety and efficacy and establishing critical quality attributes (CQAs) are essential for both innovator and biosimilar drugs.2,3 Characterization generally involves chromatography (e.g., SEC, RP, HILIC) combined with mass spectrometry (MS).

Trastuzumab received approval for medical usage in the United States in September 1998 and in the European Union in August 2000. At present, five biosimilars have gained approval in both the European Union and the United States.4,5

This application describes the utilization of a nonreducing RPLC-MS workflow for intact mass comparison of trastuzumab innovator and biosimilar. The report includes the correlation of de-charged, observed masses with theoretical masses.

A recombinant human antibody standard, SILu™Lite SigmaMAb™  (MSQC4), is utilized as a reference and assay control sample. Here, this is referred to simply as SigmaMAb™  although several other SigmaMAb™ standards are also available commercially.


Workflow for Intact Mass Analysis of Trastuzumab

Antibody reduction (optional)

Icon for optional reduction process

Chromatography

Icon for chromatography

Measurement and Analysis

Icon for measurement and analysis

A complete RPLC-MS workflow has been developed to simplify intact mass analysis of non-reduced monoclonal antibodies (mAbs) for biosimilarity assessment.

In detail, it includes:

  • Mass spectrometer calibration
  • System suitability test utilizing a recombinant human monoclonal antibody reference
  • A generic RPLC-MS method compatible with sample separation and analysis of both non-reduced and reduced monoclonal antibodies

Experimental - Sample and Reference Preparation and System Setup

Instrument Calibration

The Waters QToF Xevo G2XS mass spectrometer was calibrated in a mass range of 400 – 4000 m/z with a 10 μL/min infusion of 0.4 mg/mL of sodium iodide in water. Alternatively, calibration can be performed with a 10 μL/min infusion of 0.4 mg/mL cesium iodide, polyalanine in water prior to running the samples.

System Suitability

To evaluate performance of the entire workflow, an assay control (SigmaMAb™) was prepared and analyzed along with the samples. SigmaMAb™ reference was tested to ensure system suitability.

RPLC-MS System Setup and MS Data Analysis

RPLC-MS System Setup

The essential settings of the UHPLC-PDA chromatography system and the qToF mass spectrometer applied in the analysis of non-reduced antibodies are listed in Tables 1 and 2 below.

LC  Conditions

Instrument:

Waters™ H-Class Acquity UPLC Chromatography System

Software:

UNIFI

Column:

BIOshell™ A400 Protein C4, 100 x 2.1 mm I.D., 3.4 μm (66825-U)

Mobile phase:

[A] Water + 0.1% FA; [B] acetonitrile + 0.1% FA

Gradient:

See Table 1b

Flow:

0.2 mL/min

Column temp:

80 °C

Autosampler temp:

8 °C

Run time:

10 min

Detector:

PDA 280 nm and MS TOF (see Table 2 for conditions)

Flow divert:

1.0 –10.0 min

Loop volume:

10 μL

Injection method:

Partial loop or full loop

Injection volume:

1 μL for intact mass

Table 1.UHPLC-PDA settings.

Time (min)

%A

%B

Curve

0.00

95.0

5.0

Initial

1.00

95.0

5.0

6

3.50

40.0

60.0

6

5.00

40.0

60.0

6

5.50

5.0

95.0

6

7.50

5.0

95.0

6

7.60

95.0

5.0

6

10.00

95.0

5.0

6

Table 1b.Gradient Table for intact mass analysis

MS Conditions

 

Instrument:

Waters™ QToF Xevo G2X2 Mass Spectrometer

Software:

UNIFI

Capillary (V):

3,000

Sample cone (V)

40

Extraction cone (V):

3

Ion guide (V):

3

Desolvation temp (° C):

100

Source temp (° C):

120

Scan range (Da):

400 – 4,000

Desolvation gas (L/h):

800

Cone gas (L/h):

50

Collision energy (V):

5

Pusher (V):

930

RF Setting:

Autoprofile

Table 2.qToF-MS settings

MS Data Analysis

Data were processed using the MaxEnt1 module within the UNIFI software to generate and analyze deconvoluted (zero charged) mass spectra. In general, a summed spectrum was created from the corresponding total ion chromatogram (TIC) of the eluted intact mAb. The summed m/z spectrum was then processed by the MaxEnt1 algorithm; detailed parameters are listed in Table 3.

For glycoform analysis, data were processed using UNIFI software from Waters™. Glycoforms were matched by the software. Glycoform relative abundance data were tabulated based on peak intensities of the coeluting glycoform species.

Non-Reduced Intact Mass

m/z Range:

2,000 – 3,500

Mass range (Da):

140,000 - 160,000

Output resolution:

1 Da

Peak width model:

tof

TOF resolution:

10,000

Minimum intensity ratios, L and R (%):

30

Iterations:

12

Table 3.Deconvolution parameters.

Results

The analysis objective here, was to perform non-reduced RPLC-MS intact mass analysis on all submitted samples to compare the molecular weight of trastuzumab innovator and biosimilar drugs.

Intact SigmaMAb™ was used to determine system suitability. All mAb samples were solubilized in 100 μL water to a final concentration of 1 mg/mL.

System Suitability Test Results

SigmaMAb reference sample (1 μL) was injected on the RPLC-MS system. Figure 1 illustrates the photodiode array (280 nm) and TIC (total ion current) traces of the non-reduced antibody, while Figure 2 displays the deconvoluted mass spectrum of the SigmaMAb™ reference. The observed intact mAb glycoforms matched the common glycoform masses of MSQC4, as listed in Table 4. The measured discrepancies between the observed masses and the theoretical values for four glycoforms are all within 0.011 % mass error.

A UV chromatogram of non-reduced SigmaMAb reference standard, measured at 280 nm, displays a distinct peak at 4.72 minutes, with absorbance on the y-axis and retention time on the x-axis.
A TIC chromatogram of non-reduced SigmaMAb reference standard displays a distinct peak at 4.84 minute, with total ion current on the y-axis and retention time on the x-axis.

Figure 1. Photodiode array (280 nm, left) and TIC traces (right) of non-reduced SigmaMAb™ reference.

Summed mass spectra for non-reduced SigmaMAb reference with intensity measured in counts on the y-axis and observed mass on the x-axis
Deconvoluted mass spectra obtained for non-reduced SigmaMAb reference (MSQC4) with six distinct peaks at observed mass values of 146524, 146674, 146831, 146990, 147157, and 147311.

Figure 2. MS data for non-reduced SigmaMAb™ reference (MSQC4). Left: summed spectrum; right: deconvoluted spectrum.

Species

Molecular Formula

Theoretical Mass (Da)*

Observed Mass (Da)

Error (Da)

Error (%)

Heavy Chain/G0FG0F

C6486H10048N1716O2070S46

146658

146675

17

+0.011

Heavy Chain/G0FG1F

C6492H10058N1716O2075S46

146821

146833

12

+0.008

Heavy Chain/G1FG1F

C6498H10068N1716O2080S46

146983

146994

11

+0.007

Heavy Chain/G1FG2F

C6504H10078N1716O2085S46

147145

147156

11

+0.007

Heavy Chain/G2FG2F

C6510H10088N1716O2090S46

147307

147310

3

+0.002

G0F: GlcNAc2Man3GlcNAc2Fuc

G1F: GalGlcNAc2Man3GlcNAc2Fuc

G2F: Gal2GlcNAc2Man3GlcNAc2Fuc

*Masses based on NIST Physical Reference Data

Table 4.Observed and theoretical masses of non-reduced SigmaMAb™ reference glycoforms

Non-Reduced Sample Results

The monoclonal antibody samples were analyzed in their non-reduced form using RPLC-MS. The corresponding photodiode array (280 nm) traces, TICs, MS spectra and the deconvoluted MS spectra of trastuzumab innovator and biosimilar are shown in Figures 3 and 4. The relative quantitation results for N-glycoform distribution are shown in Figure 5. The method detects differences in the distribution of major N-glycoforms between innovator and biosimilar products. The observed masses of the non-reduced mAb correlate well with the calculated theoretical masses for all submitted samples, as shown in Table 5, and the observed mass error is 0.004% or less.

A UV chromatogram of non-reduced trastuzumab, measured at 280 nm, exhibits a distinct peak at 4.67 minutes. The innovator is represented by a red curve, while the biosimilar is depicted by a blue curve, with absorbance on the y-axis and retention time on the x-axis.
A TIC chromatogram of fully reduced trastuzumab, exhibits a distinct peak at 4.79 minutes. The innovator is represented by a red curve, while the biosimilar is depicted by a blue curve, with absorbance on the y-axis and retention time on the x-axis.

Figure 3. Comparative photodiode array (280 nm, left) and TIC traces (right) of non-reduced trastuzumab. Red: Innovator; blue: Biosimilar.

Summed up mass spectra for non-reduced trastuzumab. The innovator is represented by a red curve, while the biosimilar is depicted by a blue curve, with intensity measured in counts on the y-axis and observed mass on the x-axis.
The deconvoluted mass spectra for non-reduced trastuzumab reveals distinct peaks at mass values of 148059, 148221, 148383, and 148546 from left to right. The innovator is depicted by a red curve, and the biosimilar by a blue curve, with intensity measured in counts on the y-axis and observed mass on the x-axis.

Figure 4. Comparative MS data for non-reduced trastuzumab. Left: summed spectrum; right: deconvoluted spectrum. Red: Innovator; blue: Biosimilar

A bar graph depicting the distribution of major N-glycoforms of trastuzumab by Intact RPLC-MS. The innovator is depicted in red curve, and the biosimilar in blue, with relative abundance measured on the y-axis.

Figure 5.N-glycoform distribution of trastuzumab by Intact RPLC-MS. Red: Innovator; Blue: Biosimilar.

Species

Theoretical Mass

(Da)*

Observed Mass (Da)

Error (Da)

Error (%)

Innovator

Biosimilar

Innovator

Biosimilar

Innovator

Biosimilar

Heavy Chain/G0FG0F

148056

148061

148061

+5

+5

+0.003

+0.003

Heavy Chain/G0FG1F

148218

148222

148223

+4

+5

+0.003

+0.003

Heavy Chain/G1FG1F

148380

148384

148386

+4

+6

+0.003

+0.004

Heavy Chain/G1FG2F

148542

148546

148547

+4

+5

+0.003

+0.003

Heavy Chain/G2FG2F

148704

148707

148705

+3

+1

+0.002

+0.001

G0F: GlcNAc2Man3GlcNAc2Fuc

G1F: GalGlcNAc2Man3GlcNAc2Fuc

G2F: Gal2GlcNAc2Man3GlcNAc2Fuc

*Masses based on NIST Physical Reference Data

Table 5.Calculated and experimental masses of non-reduced trastuzumab

Conclusion

Ensuring the regulatory approval of biosimilar monoclonal antibody (mAb) drug products necessitates a meticulous and effective comparability assessment to guarantee their safety and efficacy. This application note serves as an illustration of the utilization of reversed-phase liquid chromatography-mass spectrometry (RPLC-MS) at the intact level. The primary objective revolves around assessing the comparability between the original trastuzumab drug and its biosimilar counterpart. A comprehensive workflow was devised for intact mass analysis of non-reduced monoclonal antibodies, using trastuzumab as a representative mAb, and a SILu™Lite SigmaMAb™ product, as an assay control.

A separate workflow (Application Note: Biosimilar Comparability Workflow by Middle-Up Mass Analysis Using Trastuzumab as Model Protein - Protocol for reduction and RPLC-MS analysis of a monoclonal antibody) describes an optional mAb reduction procedure which can be performed with either a partial, or full, reduction procedure. Depending on the procedure used, intra-chain disulfide bonds remain intact (partial reduction) or are broken (full reduction). Solutions of sodium iodide, or cesium iodide, were used for mass spectrometer calibration, and a system suitability test, utilizing a recombinant human monoclonal antibody reference, were developed for routine use. In addition, an RP method suitable for separation and analysis of both reduced and non-reduced mAbs was established using a wide-pore, superficially porous C4 column.

Results obtained from the non-reduced SigmaMAb™ reference indicated minimal disparities of 0.011% mass error, or less, between observed and theoretical masses for four glycoforms. For trastuzumab, both the innovator and biosimilar samples displayed a strong concurrence between observed and theoretical non-reduced mAb masses, with an observed mass error of 0.004%, or lower.

The experimental findings affirm the versatility of the workflow for analyzing non-reduced monoclonal antibody samples, delivering accurate results that enable distinct identification of multiple glycoforms.

See more applications on the Biopharmaceutical Characterization page.

References

1.
Review and Approval. 2022 Dec 13 . [Internet]. U.S. Food and Drug Administration. Available from: https://www.fda.gov/drugs/biosimilars/review-and-approval
2.
Sandra K, Vandenheede I, Sandra P. 2014. Modern chromatographic and mass spectrometric techniques for protein biopharmaceutical characterization. Journal of Chromatography A. 133581-103. https://doi.org/10.1016/j.chroma.2013.11.057
3.
Berkowitz SA, Engen JR, Mazzeo JR, Jones GB. 2012. Analytical tools for characterizing biopharmaceuticals and the implications for biosimilars. Nat Rev Drug Discov. 11(7):527-540. https://doi.org/10.1038/nrd3746
4.
Biosimilars approved in the US. Gabionline.net. [Internet]. Available from: https://www.gabionline.net/biosimilars/general/biosimilars-approved-in-the-us
5.
Biosimilars approved in Europe. Gabionline.net. [Internet]. Available from: https://www.gabionline.net/biosimilars/general/biosimilars-approved-in-europe
6.
Protocol for purification, optional reduction, and SEC-MS analysis of a monoclonal antibody. [Internet]. Sigmaaldrich.com. Available from: /US/en/technical-documents/protocol/pharmaceutical-and-biopharmaceutical-manufacturing/biologics-biosimilars-characterization/protocol-for-analysis-of-a-monoclonal-antibody