Essential Oil on SLB®-5ms

column SLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U) oven 50 °C, 3 °C/min to 250 °C (5 min) carrier gas helium, 30 cm/sec, constant sample cinnamoµm camphora essential oil diluted 1:10 in n-hexane injection 1 µL, 50:1 split inj. temp. 280 °C detector FlD, 280 °C Analysis Note Cinnamoµm Camphor, al...

GC Analysis of Peppermint Essential Oil on SUPELCOWAX® 10

column SUPELCOWAX 10, 60 m x 0.25 mm I.D., 0.25 µm (24081) oven 75 C (4 min.), 4 °C/min. to 200 °C (5 min.) carrier gas helium, 25 cm/sec @ 155 C injection 0.2 µL, 100:1 split detector FID Categories Analytical Chromatography, GC Applications, Essential Oils Featured Industry Flavors and Fragrances ...

GC Analysis of Peppermint Essential Oil on SLB®-5ms

column SLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U) oven 50 °C, 3 °C/min to 300 °C carrier gas helium, 30 cm/sec sample Peppermint oil purchased from Misitano&Stracuzzi S.p.a. (Messina, Italy) diluted 1:10 in n-hexane inj. temp. 220 °C detector FID, 300 °C detector 1 µL, 100:1 split Categories An...

GC Analysis of Bergamot Essential Oil on SLB-5ms (10 m x 0.10 mm I.D., 0.10 μm), Very Fast GC Analyis

column SLB-5ms, 10 m x 0.10 mm I.D., 0.10 µm (28465-U) oven 50 °C, 50.0 °C/min to 250 °C carrier gas hydrogen, 81.5 cm/sec, 265.3 kPa, sample 10:1 in hexane injection 0.2 µL, 400:1 split inj. temp. 300 °C detector FID, 300 °C Categories Analytical Chromatography, GC Applications, Essential Oils Feat...

GC Analysis of Peppermint Essential Oil on SLB®-IL59

column SLB-IL59, 30 m x 0.25 mm I.D., 0.20 µm (28891-U) oven 50 °C, 3 °C/min to 300 °C carrier gas helium, 30 cm/sec sample Peppermint oil purchased from Misitano&Stracuzzi S.p.a. (Messina, Italy) diluted 1:10 in n-hexane injection 1 µL, 100:1 split inj. temp. 220 °C detector FID, 300 °C Categories ...

GC Analysis of Bergamot Essential Oil on SLB-5ms (10 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

column SLB-5ms, 10 m x 0.10 mm I.D., 0.10 µm (28465-U) oven 50 °C, 14.0 °C/min to 250 °C carrier gas hydrogen, 57.0 cm/sec, 184.2 kPa sample bergamot oil, diluted 10:1 in hexane injection 0.4 µL, 600:1 split liner 0.75 mm ID inj. temp. 300 °C detector FID, 300 °C Categories Analytical Chromatography...

GC Analysis of Lemon Essential Oil on SLB®-5ms (10 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

column SLB-5ms, 10 m x 0.10 mm I.D., 0.10 µm (28465-U) oven 40 °C, 50 °C/min to 320 °C carrier gas hydrogen, 81.5 cm/sec constant sample lemon essential oil in hexane injection 0.4 µL in the split mode (300:1) inj. temp. 320 °C detector FID, 320 °C Categories Analytical Chromatography, GC Applicatio...

GC Analysis of Petitgrain Essential Oil on SLB®-5ms, 50 °C Initial Oven Temp., 30 cm/s

column F.S. CAP. SLB-5ms 30 m x 0.25 mm I.D., 0.25 µm (58471-U) oven 50 °C, 3 °C/min to 250 °C (10 min) carrier gas helium, 30.0 cm/s, 99.5 kPa sample (oil obtained by steam distillation of the leaves and twigs, diluted) injection 1.0 µL, 100:1 split inj. temp. 250 °C detector FID, 280 °C Analysis N...

GC Analysis of Spearmint Essential Oil on SLB®-IL59, 80 °C Initial Oven Temp.

column SLB-IL59, 30 m x 0.25 mm I.D., 0.20 µm (28891-U) oven 80 °C, 3 °C/min to 280 carrier gas helium, 40 cm/sec sample spearmint essential oil diluted 1:10 in hexane injection 1 µL100:1 split inj. temp. 250 °C detector FID, 305 °C Categories Analytical Chromatography, GC Applications, Essential Oi...

GC Analysis of Neroli Essential Oil on SLB®-5ms

column F.S. CAP. SLB-5ms 30 m x 0.25 mm I.D., 0.25mum (28471-U) oven 50 °C, 3 °C/min to 250 °C (10 min) carrier gas helium, 30.0 cm/sec, 99.5 kPa sample oil obtained from the steam distillation or hydrodistillation of the flowers, diluted injection 1.0 µL,100:1 split (A032) inj. temp. 250 °C detecto...

GC Analysis of Petitgrain Essential Oil on SUPELCOWAX 10, 50 °C Initial Oven Temp.

column SUPELCOWAX 10, 30 m x 0.25 mm I.D., 0.25 µm (24079) oven 50 °C (0 min), then 3 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 260 °C Ca...

GC Analysis of Petitgrain Essential Oil on SLB®-IL60, 50 °C Initial Oven Temp.

column SLB-IL60, 30 m x 0.25 mm I.D., 0.25 µm (29505-U) oven 50 °C (0 min), then 3 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 260 °C Categ...

GC Analysis of Petitgrain Essential Oil on SLB®-5ms, 50 °C Initial Oven Temp.

column SLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U) oven 50 °C (0 min), then 3 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Catego...

GC Analysis of Patchouli Essential Oil on SLB®-5ms

column SLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U) oven 50 °C (0.5 min), then 5 °C/min to 300 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Patchouli Oil, neat (W283800) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 280 °C detector FID, 310 °C Categories Anal...

GC Analysis of Patchouli Essential Oil on SLB®-IL60

column SLB-IL60, 30 m x 0.25 mm I.D., 0.25 µm (29505-U) oven 50 °C (0 min), then 5 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Patchouli Oil, neat (W283800) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Categories Analy...

GC Analysis of Lemon Essential Oil on SLB®-IL59, 3 °C/min Oven Ramp

column SLB-IL59, 30 m x 0.25 mm I.D., 0.2 µm (28891-U) oven 50 °C, 3 °C/min to 300 °C carrier gas helium, 30 cm/sec, constant detector FID Analysis Note The volatile fraction of essential oils represents a quite complex matrix, being composed of a large variety of compounds, mainly terpenes, sesquit...

GC Analysis of Petitgrain Essential Oil on Equity®-1

column Equity-1, 30 m x 0.25 mm I.D., 0.25 µm (28046-U) oven 75 C (8 min), then 4 °C/min to 200 °C (15 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL, 100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Cate...

GC Analysis of Lemon Essential Oil (Showing Camphene/Δ-3-Carene Ratio) on SLB®-5ms (10 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

column SLB-5ms, 10 m x 0.10 mm I.D., 0.10 µm (28465-U) oven 40 °C, 30 °C/min to 85 °C, 80 °C/min to 320 °C carrier gas hydrogen, 70 cm/sec constant sample lemon essential oil in hexane; the value for delta-3-carene/camphene must be <0.140 to be considered pure lemon essential oil injection 0.4 µL in...

GC Analysis of Lemon Essential Oil Contaminated with 5%Sweet Orange Essential Oil (Showing Camphene/Δ-3-Carene Ratio) on SLB®-5ms (10 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

column SLB-5ms, 10 m x 0.10 mm I.D., 0.10 µm (28465-U) oven 40 °C, 30 °C/min to 85 °C, 80 °C/min to 320 °C carrier gas hydrogen, 70 cm/sec constant sample lemon essential oil + 5% sweet orange essential oil in hexane; the delta-3-carene/camphene value is 0.275, which exceeds the maxiµm value of 0.14...

GC Analysis of Spearmint Essential Oil on SLB®-IL59

column SLB-IL59, 30 m x 0.25 mm I.D., 0.2 µm (28891-U) oven 50 °C, 3 °C/min to 300 °C carrier gas helium, 30 cm/sec, constant detector FID Analysis Note The volatile fraction of essential oils represents a quite complex matrix, being composed of a large variety of compounds, mainly terpenes, sesquit...

GC Analysis of Petitgrain Essential Oil on SUPELCOWAX® 10

column SUPELCOWAX 10, 30 m x 0.25 mm I.D., 0.25 µm (24079) oven 75 C (4 min), then 4 °C/min to 200 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Cat...

GC Analysis of Petitgrain Essential Oil on SLB®-IL60

column SLB-IL60, 30 m x 0.25 mm I.D., 0.25 µm (29505-U) oven 75 C (4 min), then 4 °C/min to 200 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Catego...

GC Analysis of Petitgrain Essential Oil on SLB®-5ms

column SLB-5ms, 30 m x 0.25 mm I.D., 0.25 µm (28471-U) oven 75 °C (8 min), then 4 °C/min to 200 °C (15 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Catego...

GC Analysis of Patchouli Essential Oil on Equity®-1

column Equity-1, 30 m x 0.25 mm I.D., 0.25 µm (28046-U) oven 50 °C (0.5 min), then 5 °C/min to 300 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Patchouli Oil, neat (W283800) injection 0.1 µL, 100:1 split liner 4 mm I.D. cup injector inj. temp. 280 °C detector FID, 310 °C Categories A...

GC Analysis of Patchouli Essential Oil on SUPELCOWAX® 10

column SUPELCOWAX 10, 30 m x 0.25 mm I.D., 0.25 µm (24079) oven 50 °C (0 min), then 5 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Patchouli Oil, neat (W283800) injection 0.1 µL100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 °C Categories An...

GC Analysis of Lemon Essential Oil on SLB®-IL59

column SLB-IL59, 30 m x 0.25 mm I.D., 0.20 µm (28891-U) oven 50 °C, 5 °C/min to 300 °C carrier gas helium, 40 cm/sec sample lemon essential oil diluted 1:10 in hexane injection 1 µL100:1 split inj. temp. 250 °C detector FID, 305 °C Categories Analytical Chromatography, GC Applications, Essential Oil...

GC Analysis of Spearmint Essential Oil on SUPELCOWAX® 10

column SUPELCOWAX 10, 30 m x 0.25 mm I.D., 0.25 µm (24079) oven 75 °C (4 min), 4 °C/min to 200 °C (5 min) carrier gas helium, 25 cm/sec @ 155 °C injection 0.2 µL, 100:1 split detector FID Categories Analytical Chromatography, GC Applications, Essential Oils Featured Industry Flavors and Fragrances L...

GC Analysis of Petitgrain Essential Oil on Equity®-1, 50 °C Initial Oven Temp.

column Equity-1, 30 m x 0.25 mm I.D., 0.25 µm (28046-U) oven 50 °C (0 min), then 3 °C/min to 250 °C (10 min) carrier gas helium, 25cm/ sec set at 110C sample Petitgrain Oil, Paraguay, neat (W523607) injection 0.1 µL, 100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 260 °C Cat...

GC Analysis of Native Spearmint Essential Oil on Equity®-1

column Equity-1, 30 m x 0.25 mm I.D., 0.25 µm (28046-U) oven 75 °C (8 min), then 4 °C/min to 200 °C (15 min) carrier gas helium, 25cm/ sec set at 110C sample Native Spearmint Oil, (A.M. Todd Sample), neat injection 0.1 µL, 100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, 250 ...

GC Analysis of Willamette Peppermint Essential Oil on Equity®-1

column Equity-1, 30 m x 0.25 mm I.D., 0.25 µm (28046-U) oven 75 C (8 min), then 4 °C/min to 200 °C (15 min) carrier gas helium, 25cm/ sec set at 110C sample Willamette Peppermint Oil, (A.M. Todd Sample), neat injection 0.1 µL, 100:1 split liner 4 mm I.D. cup injector inj. temp. 250 °C detector FID, ...