GC Analysis of Aromatics in Reformulated Gasoline (Ethanol at 20%) on SLB®-IL111
Conditions
column | SLB-IL111, 60 m x 0.25 mm I.D., 0.20 µm (28928-U) |
oven | 50 °C (3 min), 15 °C/min to 265 °C (5 min) |
carrier gas | helium, 30 cm/sec |
sample | Premium unleaded gasoline, plus ethanol at 20% (v/v) and two internals each at 4% (v/v) |
injection | 0.5 µL, 200:1 split |
liner | 4 mm I.D., split type, single taper wool packed FocusLiner™ design |
inj. temp. | 250 °C |
detector | FID, 275 °C |
Description
Categories | Analytical Chromatography, GC Applications, Aromatics, Hydrocarbons |
Featured Industry | Petroleum |
Legal Information | FocusLiner is a trademark of SGE Analytical Science Pty Ltd., SLB is a registered trademark of Sigma-Aldrich Co. LLC |
Other Notes | Complementary products: Syringes Vials Gas Purification and Gas Management Inlet Liners, Glass Wool, and Specialized Hand Tools Septa and Specialized Hand Tools Column Ferrules, Nuts, and Specialized Hand Tools |
suitability | application for GC |