Attention:

We are currently upgrading the certain systems that support this website.  You may experience some functionality loss, though we expect the site to be fully functional by 9:00PM CT (US), February 24th.

We apologize for any inconvenience and appreciate your patience during this improvement process.Please note that you still have telephone and email access to our local offices. We apologize for any inconvenience.

HPLC Analysis of Terbinafine HCl Cream on Purospher(R) Star Phenyl (5mum) 150 x 3 mm

HPLC Analysis of Terbinafine HCl Cream on Purospher® Star Phenyl (5mum) 150 x 3 mm

Conditions

sample preparation Dissolve appropriate amount of Terbinafine Hydrochloride Cream in isopropanol to obtain a 0.2 mg/mL (200 ppm) of solution, filter prior to analysis. Light degradation solution expose 1mg/mL of Terbinafine Hydrochloride in 50% ACN solution at 254 nm of UV light for 1h. Inject the solution to HPLC system, 3 major related impurity can be found with relative retention time (RRT) from 0.8 to 1.2 (Generally 0.87, 0.95,1.1).
column Purospher® STAR RP-18 endcapped (5 mm) 150 x 3 mm (1.50626.0001)
column temp. 25 °C
mobile phase Mobile Phase A - [A] 2.0 mL/L of triethylamine. Adjust pH 7.5 with glacial acetic acid; [B] methanol; [C] acetonitrile; (30:42:28, A:B:C). Mobile Phase B - [A] 2.0 mL/L of triethylamine. Adjust pH 7.5 with glacial acetic acid; [B] methanol; [C] acetonitrile; (5:57:38, A:B:C)
gradient Time (min), A (%), B (%); 0, 100, 0; 4, 100, 0; 25, 0, 100; 30, 0, 100; 31, 100, 0; 38, 100, 0;
flow rate 0.8 mL/min
pressure 2407 psi (166 Bar)
sample Dilute appropriate amount of standard with isopropanol to obtain a 0.2 mg/mL (200 ppm) solution.
injection 20 μL
detector 280 nm cell: 1 μL/10 mm

Description

Featured Industry Pharmaceutical (small molecule)
Legal Information Purospher is a registered trademark of Merck KGaA, Darmstadt, Germany
suitability application for HPLC