Analytical Method: Nickel in plating baths

Instruments and Reagents:

Cat.No. 173016  Spectroquant® VIS Spectrophotometer Prove 100 or

Cat.No. 173017  Spectroquant® UV/VIS Spectrophotometer Prove 300 or

Cat.No. 173018  Spectroquant® UV/VIS Spectrophotometer Prove 600 or

Cat.No. 109748  Spectroquant® Photometer NOVA 30 or

Cat.No. 109751  Spectroquant® Photometer NOVA 60 or

Cat.No. 173632  Spectroquant® Colorimeter Move 100

Cat.No. 114946  Rectangular cells 10 mm and/or

Cat.No. 114947  Rectangular cells 20 mm and/or

Cat.No. 114944  Rectangular cells 50 mm
Cat.No. 114554  Spectroquant® Nickel Cell Test
Cat.No. 114785  Spectroquant® Nickel Test
Cat.No. 116754  Water for analysis
 
Sample preparation:
In a 500-ml standard volumetric flask place 2.5 ml of bath solution, which has been cooled to room temperature, fill up to volume with water for analysis and mix well. Take from this solution a further 5 ml and place in a second 500-ml standard volumetric flask and again fill up with water for analysis and mix.
Determine readily the nickel content in solutions diluted in this fashion.
 
Analysis:
Determine with the above mentioned test kits
 
Calculation:
Nickel content in g/l = analysis value in mg/l Ni x 20