material
316L stainless steel holder (EN 1.4435; Locking Nut and Tool: Stainless steel SS2346, ASTM 303), EPDM rubber O-ring, polyphenylene sulfide (PPS) magazine (blue plastic portion)
needle size
2.0 mm
sterility
non-sterile
product line
NovaSeptum®
feature
autoclavable
packaging
pkg of 1 unit
parameter
4-150 °C temp. range (39-302 °F), 6 bar max. pressure (87 psi)
fitting
50.5 mm TC inlet, 50.5 mm TC outlet
compatibility
for use with NovaSeptum® Cell Culture Sampling Units
storage temp.
room temp
General description
The NovaSeptum® Cell Culture Sampling TC Holder is a stainless steel sampling holder designed to connect NovaSeptum® sterile sampling units to process vessels and piping through a standard Tri-Clamp (TC) connection. For an optimized aseptic setup, use the holder together with a NovAseptic® connector.
Designed for use with the NovaSeptum® sterile, closed sampling system, the holder enables aseptic fluid sampling from bioreactors, tanks, and process lines while helping maintain process sterility and sample integrity.
Configured for Cell Culture Sampling (CCS) with a 2.0 mm needle, the holder features seven sampling ports, allowing sequential sample collection from the same process connection point. The larger needle diameter supports higher flow rates while helping reduce shear stress, making the holder particularly suitable for sampling high-viscosity cell culture fluids. The holder is compatible with the full range of NovaSeptum® Cell Culture Sampling units, including High-Purity bags, Accurate Volume syringes, bottles, and transfer units.
Designed for use with the NovaSeptum® sterile, closed sampling system, the holder enables aseptic fluid sampling from bioreactors, tanks, and process lines while helping maintain process sterility and sample integrity.
Configured for Cell Culture Sampling (CCS) with a 2.0 mm needle, the holder features seven sampling ports, allowing sequential sample collection from the same process connection point. The larger needle diameter supports higher flow rates while helping reduce shear stress, making the holder particularly suitable for sampling high-viscosity cell culture fluids. The holder is compatible with the full range of NovaSeptum® Cell Culture Sampling units, including High-Purity bags, Accurate Volume syringes, bottles, and transfer units.
Application
- Sterile and aseptic sampling in biopharmaceutical upstream and downstream manufacturing
- Multi-point in-process sampling from bioreactors, mixing vessels, and storage tanks with TC 50.5 mm connections
- Sterility testing, bioburden testing, and endotoxin testing sample collection
- Chemical analysis and pH analysis in aseptic processes
- Multi-use process environments requiring closed, contamination-free sampling
- Processes where product hold-up minimization and sample representativeness are critical
Features and Benefits
- 7-port TC 50.5 mm (1.5 in.) Tri-Clamp connection: Integrates directly into standard TC 50.5 mm process fittings; no additional adapter required
- 2 mm needle - Cell Culture Sampling configuration: Enables higher flow rates and reduced shear stress, making it suited for high-viscosity cell culture fluid sampling while remaining compatible with a wide range of sample types and viscosities.
- 316L stainless steel (EN 1.4435) construction: Meets pharmaceutical-grade material requirements; fully compatible with standard CIP/SIP regimes
- Electropolished + glass-blasted surface finish (Ra ≤ 0.5 μm): Minimizes microbial harborage sites, supports cleaning validation
- Compatible with NovAseptic® NAC connectors: Provides installation flexibility across global process equipment standards
- Autoclavable and SIP compatible: Supports standard sterilization workflows without requiring replacement of the holder hardware
- Design pressure: 6 bar | Design temperature: 4°C to +150°C: Suitable for a wide range of process conditions
- Certificate of quality included: Supports GMP documentation requirements and lot traceability
Packaging
All open ends on the NovaSeptum® holders are protected with plastic covers. The holder is pre-assembled and packed in a closed box to reduce exposure.
Legal Information
NOVASEPTUM is a registered trademark of Merck KGaA, Darmstadt, Germany
NOVASEPTIC is a registered trademark of Merck KGaA, Darmstadt, Germany
Still not finding the right product?
Explore all of our products under NovaSeptum® Cell Culture Sampling TC Holder
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.