We have not studied stacking Swinnex filters in series one after the other for double and separate filtration. One could attempt this but may need a stronger vacuum source to pull samples through both or a stronger pushing source such as a syringe. Depending on the sample matrix with particle type and concentration, one could also try stacking two filters on top of each other in one device. Both options have not been attempted and there is a higher element of risk of this functioning on many initial studies.
Select a Size
About This Item
Skip To
material
polypropylene device
silicone gasket
silicone seal
Quality Level
Agency
suitable for EPA 200.7
suitable for EPA 200.8
suitable for EPA 300
manufacturer/tradename
Swinnex®
parameter
5.2 bar max. inlet pressure (75 psi)
technique(s)
pressure filtration: suitable
L
5.4 cm
diam.
5.8 cm (2.3 in.)
filter diam.
47 mm
filtration area
13.8 cm2
prefilter diam.
42 mm , thick depth prefilter
47 mm , membrane prefilter
fitting
1/4 in. NPT inlet connection thread(s)
1/4 in. male NPT inlet fitting (with female Luer slip interior)
female Luer outlet slip
application(s)
filtration
compatibility
for use with Sterifil Aseptic System
shipped in
ambient
Related Categories
1 of 4
This Item | SX0002500 | SX0001300 | XX3001200 |
|---|---|---|---|
| manufacturer/tradename Swinnex® | manufacturer/tradename Swinnex® | manufacturer/tradename Swinnex® | manufacturer/tradename - |
| material polypropylene device, silicone seal, silicone gasket | material polypropylene device, silicone gasket, silicone seal | material polypropylene device, silicone gasket, silicone seal | material PTFE O-ring, PTFE gasket, PTFE seal, stainless steel housing, stainless steel screen |
| agency suitable for EPA 200.7 | agency suitable for ASTM® 7968, suitable for ASTM® 7979, suitable for EPA 8327 | agency suitable for ASTM® 7968, suitable for ASTM® 7979, suitable for EPA 8327 | agency - |
| parameter 5.2 bar max. inlet pressure (75 psi) | parameter 121 °C max. temp., 3.5 bar max. inlet pressure (50 psi) | parameter 121 °C max. temp., 3.5 bar max. inlet pressure (50 psi) | parameter 6.9 bar max. inlet pressure (100 psi) |
| technique(s) pressure filtration: suitable | technique(s) pressure filtration: suitable | technique(s) pressure filtration: suitable | technique(s) pressure filtration: suitable |
| L 5.4 cm | L 3.7 cm | L 3.5 cm | L 3.3 cm |
General description
Features & Benefits:
•Available in three sizes for a range of process volumes
•Autoclavable with filter in place for sterile filtration
SX0001300 Swinnex Filter Holder, 13 mm
SX0002500 Swinnex Filter Holder, 25 mm
SX0004700 Swinnex Filter Holder, 47 mm
Replacement Parts
SX0001301 Silicone Gaskets for 13 mm Swinnex Filter Holder
SX0002501 Silicone Gaskets for 25 mm Swinnex Filter Holder
XX2702510 Support Screen, 25 mm, polypropylene
XX1104715 Support Screen, 47 mm, polypropylene
XX1104707 O-ring (5-329) for Swinnex 47 mm Filter Holder, silicone
SX0004701 Base O-ring for Swinnex 47 mm Filter Holder, silicone
SX00047RK Replacement Parts Kit for Swinnex 47 mm Filter Holder
Biochem/physiol Actions
Preparation Note
Autoclavable with filter in place
Legal Information
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Protocols
This is a User Guide Dispensing Pressure Vessel Fitting Kit from Millipore.
Related Content
This page is intended to make it easier to find the consumables you need based on the analytical method you’re using. Methods included on this page come from the EPA, Standard Methods and ASTM.
-
Can you stack these filter holders so liquid passes from a syringe, through the first filter holder, then a second filter holder and then out the other end?
1 answer-
Helpful?
-
-
Is it necessary to use two different O-rings for the Swinnex 47 (SX0004700)?
1 answer-
Yes, two different o-rings are required for the Swinnex 47 mm filter holder (SX0004700). One is for the inlet, which is a cone-shaped component, and the other is for the base, which has a flat surface to hold the filter.
The XX1104707 o-ring seals the filter membrane in place and fits in the groove on the top (inlet) part, while the SX0004701 is the base o-ring and goes on the base (outlet).
Helpful?
-
-
What type of syringe is compatible with the SX0004700? Is an adapter required for this syringe?
1 answer-
It is recommended to use luer slip syringe (not luer lok) for the Swinnex 47 SX0004700, as it fits inside the inlet. No additional adapter is required. Additionally, the Swinnex 47 has a ¼ NPTM inlet, which includes the female luer slip interior.
Here are examples of syringes that are compatible with the Swinnex 47 and their sample holding capacities:
Z683590-100EA: Polypropylene LUER SLIP syringe, 10 ml, pack of 100
XX1102012: 20mL Luer slip polypropylene syringe, pack of 12
XX1105005: 50mL Luer slip polypropylene syringe, pack of 5Helpful?
-
-
Good day, I see in packaging it states 8pkg - just to be sure, does that mean that the prize is for a package of 8 holders?
1 answer-
Yes, this is a package of 8 Swinnex 47mm Filter Holders.