Direkt zum Inhalt
Merck

LC-MS Analysis of Pain Management Opiates using Ascentis® Express Biphenyl

LC-MS Analysis of Pain Management Opiates using Ascentis® Express Biphenyl application for LC-MS

Materialien

Ähnliches Produkt

Produkt-Nr.
Beschreibung
Preisangaben

Kodein -Lösung

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

(±)-Methadon -Lösung

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Hydromorphon -Lösung

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Buprenorphin -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Fentanyl -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Hydrocodon -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Morphin -Lösung

100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Meperidin -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Naloxone solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Naltrexone solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Oxycodon -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Oxymorphon -Lösung

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

cis-Tramadol hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Ascentis® Express Biphenyl Column

2.7 μm particle size, L × I.D. 5 cm × 2.1 mm

CONDITIONS

column

Ascentis Express Biphenyl, 5 cm x 2.1mm, 2.7 μm particles (64057-U)

mobile phase

[A] water with 0.1% formic acid; [B] acetonitrile with 0.1% formic acid;

gradient

10 to 20% B in 1.25 min; to 95% B in 3.75 min

flow rate

0.3 mL/min

pressure

2990 psi (206 bar)

column temp.

30 °C

detector

MS-TOF, ESI+

injection

2 μL

sample

50-500 ng/mL in 99:1, water:methanol

Beschreibung

Hinweis zur Analyse

This application demonstrates the separation of 13 pain managment opiates on the Ascentis Express Biphenyl using an LC-TOF mass spectrometer. This application demonstrates the scalability of this method to 50 x 2.1 mm column geometry using the 2.7 um biphenyl phase. The method provides baseline resolution of all isobaric sets of interest.

Rechtliche Hinweise