This kit has not been validated for mouse beta interleukin 1 (IL-1b). In cross-reactivity tests, only four mouse samples were tested, and three out of four were not detected, indicating signals above background levels. Non-detectable results could be due to either a lack of cross-reactivity or sample concentrations being below the detection limit.
About This Item
Skip To
Product Name
MILLIPLEX® Human High Sensitivity T Cell Panel, Configurable 96-well Human High Sensitivity T Cell 21-Plex Panel
species reactivity
human
Quality Level
packaging
pkg of 1 ea
manufacturer/tradename
Milliplex®
assay range
accuracy: 98-107%
sensitivity: 0.11-8.17 pg/mL
(Overnight Incubation minDC)
standard curve range: 0.18-750 pg/mL
(IL-6)
standard curve range: 0.24-1,000 pg/mL
(IL-13)
standard curve range: 0.24-1,000 pg/mL
(IL-21)
standard curve range: 0.31-1,250 pg/mL
(IL-8)
standard curve range: 0.31-1,250 pg/mL
(MIP-1α)
standard curve range: 0.37-1,500 pg/mL
(IL-7)
standard curve range: 0.45-1,750 pg/mL
(TNFα)
standard curve range: 0.49-2,000 pg/mL
(IL-1β)
standard curve range: 0.49-2,000 pg/mL
(IL-12 (p70))
standard curve range: 0.49-2,000 pg/mL
(IL-2)
standard curve range: 0.49-2,000 pg/mL
(IL-5)
standard curve range: 0.61-2,500 pg/mL
(IFNγ)
standard curve range: 0.61-2,500 pg/mL
(MIP-3α)
standard curve range: 0.73-3,000 pg/mL
(IL-17A)
standard curve range: 0.92-3,750 pg/mL
(MIP-1β)
standard curve range: 1.22-5,000 pg/mL
(GM-CSF)
standard curve range: 1.46-6,000 pg/mL
(IL-10)
standard curve range: 1.46-6,000 pg/mL
(ITAC)
standard curve range: 1.83-7,500 pg/mL
(IL-4)
standard curve range: 18.3-75,000 pg/mL
(Fractalkine)
standard curve range: 7.93-32,500 pg/mL
(IL-23)
inter-assay cv: <15%
intra-assay cv: <10%
technique(s)
multiplexing: suitable
input
serum
cell culture supernatant
detection method
fluorometric (Luminex® xMAP® technology)
shipped in
wet ice
storage temp.
2-8°C
1 of 4
This Item | MHSTCMAG-70K | MHSTCMAG-70KPMX | HCYTA-60K-PX48 |
|---|---|---|---|
| species reactivity human | species reactivity mouse | species reactivity mouse | species reactivity human |
| assay range accuracy: 98-107%, standard curve range: 0.18-750 pg/mL | assay range accuracy: 82-102%, sensitivity: 0.26-11.14 pg/mL, standard curve range: 0.10-400 pg/mL | assay range accuracy: 82-102%, sensitivity: 0.26-11.14 pg/mL, standard curve range: 0.10-400 pg/mL | assay range - |
| manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® | manufacturer/tradename Milliplex® |
| Quality Level 200 | Quality Level 200 | Quality Level 200 | Quality Level 200 |
| detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex xMAP) | detection method fluorometric (Luminex® xMAP®) |
| shipped in wet ice | shipped in ambient | shipped in ambient | shipped in - |
Legal Information
Application
Disclaimer
Features and Benefits
General description
related product
signalword
Danger
Hazard Classifications
Acute Tox. 4 Dermal - Acute Tox. 4 Inhalation - Acute Tox. 4 Oral - Aquatic Chronic 2 - Eye Dam. 1 - Skin Sens. 1 - STOT RE 2
target_organs
Respiratory Tract
Storage Class
10 - Combustible liquids
Certificates of Analysis (COA)
Search for Certificates of Analysis (COA) by entering the products Lot/Batch Number. Lot and Batch Numbers can be found on a product’s label following the words ‘Lot’ or ‘Batch’.
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
Related Content
Immunology multiplex assays, like MILLIPLEX® assays, allow researchers to investigate more immune biomarkers simultaneously, saving time and resources.
-
Hello, I have a question about this kit that I recently bought, and my question is whether this kit also detects beta interleukin in mouse.
1 answer-
Helpful?
-
-
Could you provide me with more information about the T cell cytokine assay, including CV values and detection limits? The current information I have is that serum concentrations of 13 human cytokines (IL-1β, IL-2, IL-4, IL-5, IL-6, IL-7, IL-8, IL-10, IL-12p70, IL-13, GM-CSF, IFN-γ, and TNF-α) were quantified (pg/ml) in the serum of 52 matched cord blood and maternal venous blood samples using the human 13-plex magnetic bead assay (HCYTOMAG-60K). The assay protocol followed the manufacturer's guidelines, and standards and controls were included in each run. I also cc-ed Jennifer as she was involved in my work and may have previous information about technical reports.
1 answer-
The validation data and assay characteristics are provided in the back of each assay protocol.