Wybierz wielkość
Informacje o tej pozycji
Nazwa produktu
Lynx® S2S, 3/8" Lynx® S2S Connector Set, 3/8" hose barb male inlet and 3/8" hose barb female outlet
material
polysulfone (21 CFR 177.1655)
silicone O-ring (21 CFR 177.2600)
silicone gasket (21 CFR 177.2600)
sterility
non-sterile
product line
Lynx®
parameter
27.3 L/min flow rate at 1.38 bar (20 psi) and 25 °C (77°F)
4-40 °C operating temperature (39-104 °F)
4.1 bar max. pressure (60 psi) (operating pressure)
fitting
3/8 in. inlet hose barb for 0.25 in (7.1 mm) I.D.
3/8 in. outlet hose barb for 0.25 in (7.1 mm) I.D.
Quality Level
Application
By design, a sterile connector should protect the fluid path by never exposing it to the open environment. The Lynx® S2S connector employs a solid plug, not a membrane or peel strip, which creates a gap for ingress of contamination during the connection process. The Lynx® S2S design provides high-pressure sealing and a flow path that remains completely closed, minimizing the risk of bacterial intrusion. It will secure the fluid path at a sterility assurance level (SAL) of 10-6 under the most rigorous conditions. Once assembled, the two components cannot be disassembled, reducing the risk of sterility breach. The male-female connection design adds safety to the process, limiting the risks of wrong connection in critical steps. Lynx® S2S: the only plug style device with a SAL of 10-6
General description
The SSCSHB3A01 Lynx S2S connector set is constructed of a female coupling and a male coupling. With these two components assembled, the sterile fluid path is enabled. Once assembled, the two components cannot be disassembled, reducing the risk of sterility breach.
This Lynx S2S unit has a 3/8 inch. diameter and a hose barb inlet/outlet connection.
Safe, Disposable Connectors
The Lynx S2S connector is disposable and intended for single-use, one-time actuation. There are no parts to clean-in-place, and the risk of contamination is reduced.
Sterility Assurance During Actuation
The Lynx S2S connector is 100% air-integrity tested in manufacturing. Connector sets were tested utilizing an aerosolized bacterial challenge method with 106 B. diminuta and direct bacterial soiling method with B. diminuta, assuring a sterile connection can be made in non-classified areas.
Preparation Note
Gamma compatible up to 45 kGy; Autoclave once for 30 minutes @ 130 °C. Do not steam in place.
Legal Information
Certyfikaty analizy (CoA)
Poszukaj Certyfikaty analizy (CoA), wpisując numer partii/serii produktów. Numery serii i partii można znaleźć na etykiecie produktu po słowach „seria” lub „partia”.
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
Nasz zespół naukowców ma doświadczenie we wszystkich obszarach badań, w tym w naukach przyrodniczych, materiałoznawstwie, syntezie chemicznej, chromatografii, analityce i wielu innych dziedzinach.
Skontaktuj się z zespołem ds. pomocy technicznej