跳转至内容
Merck

HPLC Analysis of Aflatoxins in Peanut Paste on Ascentis® Express C18 following SPE using Supel Tox AflaZea

HPLC Analysis of Aflatoxins in Peanut Paste on Ascentis® Express C18 following SPE using Supel™ Tox AflaZea application for HPLC, application for SPE

材料

相關產品

产品编号
说明
价格

黄曲霉毒素混合物 4 溶液

analytical standard

Ascentis® Express C18, 2.7 μm HPLC 色谱柱

L × I.D. 15 cm × 2.1 mm, pk of 1 ea

附屬品

产品编号
说明
价格

微型反应容器,3mL,黑色酚醛孔盖,12件装

volume 5 mL, clear glass vial, PTFE/red rubber septum, O.D. × H 20 mm × 62 mm, thread for 20-400

Visiprep SPE 真空固相萃取装置

DL (Disposable Liner), 12-port model

CONDITIONS

sample preparation

Solid Phase Extraction

sample/matrix

raw peanut paste

SPE tube/cartridge

Supel Tox AflaZea, 6 mL (55314-U)

extraction process

suspend 25 g raw peanut paste in 100 mL acetonitrile:water (84:16) extraction solution

sample addition

directly load 2 mL extract onto SPE tube

elution

pull sample through the SPE tube by vacuum at a rate of 8 mL/min

eluate post-treatment

combine 200 μL eluate and 300 μL of water:trifluoroacetic acid:acetic acid (70:20:10 ) derivatization reagent, (heat at 65°C for 25 min, add 500 μL of water to derivatized sample and vortex 30 sec)

column

Ascentis Express C18, 15 cm x 2.1 mm I.D., 2.7 μm particles (53825-U)

mobile phase

[A} acetonitrile; {B} water; (20:80, A:B)

flow rate

0.2 mL/min

column temp.

35 °C

detector

fluorescence, ex 365 nm, em 450 nm

injection

20 μL

sample

B1, G1 (8 ppb), B2, G2 (2 ppb)

说明

分析報告

Aflatoxins are carcinogenic metabolites produced by fungi that infect major agricultural commodites. The presence and concentrations of aflatoxins is widely regulated. Here, a cleanup method using interference removal SPE followed by analysis by HPLC with fluorescence detection is used to analyze derivatized samples extracted from raw peanut paste. The HPLC method was optimized for fluorescence detection and quantification of aflatoxins at < 10 ppb concentrations.

其他說明

Sample pretreatment: suspend 25 g raw peanut paste in 100 mL acetonitrile:water (84:16) extraction solution

法律資訊

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC