For Qualitative Assays, the Analysis Type is "None" in xPonent. There are no units to be entered. Median values are used for the data comparison between treatment groups.
Über diesen Artikel
product line
Milliplex®
species reactivity
human
technique(s)
multiplexing: suitable
detection method
fluorometric (Luminex® xMAP®)
storage temp.
2-8°C
General description
Application
Features and Benefits
Packaging
Preparation Note
Legal Information
Disclaimer
signalword
Warning
hcodes
Hazard Classifications
Eye Irrit. 2
Lagerklasse
10 - Combustible liquids
wgk
WGK 2
Hier finden Sie alle aktuellen Versionen:
Besitzen Sie dieses Produkt bereits?
In der Dokumentenbibliothek finden Sie die Dokumentation zu den Produkten, die Sie kürzlich erworben haben.
-
We are running the SARS-CoV-2 IgG, IgA, and IgM kits for the first time on the MagPix. What analysis settings should be used since the kit does not include a standard? Are there any units that should be entered?
1 answer-
Helpful?
-
-
Could you provide information on the subtypes detected by the SARS-CoV-2 IgG panel?
1 answer-
The antibody was generated in goats and is probably designed to detect total IgG. It does not target a specific IgG subunit.
Helpful?
-
-
What is the method for determining positive samples?
1 answer-
To determine positive samples, it is necessary to run a number of negative control, pre-COVID, samples in the assay and obtain an average MFI for each antigen to compare to the unknown samples. It's important to note that the positive and negative controls do not play a role in determining whether customer samples are positive or negative for COVID.
Helpful?
-
-
Have the Milliplex Milliplex Human SARS-CoV-2 assays been tested on species other than humans?
1 answer-
The Human SARS-CoV-2 kits were not tested with cross-species samples, so it is unknown whether they will detect samples from other species. There are no references available for using the kits with other species.
Helpful?
-