The Stericup S2GPU02RE is not DNAse or RNAse free.
Seleccione un Tamaño
| Talla/SKU | Disponibilidad | Precio |
|---|
Acerca de este artículo
Quality Segment
material
polyethersulfone membrane, polystyrene bottle, polystyrene funnel(s), polystyrene housing
sterility
sterile
product line
Stericup®
feature
hydrophilic
packaging
pack of 12 ea
parameter
45 °C max. temp.
description
2.4 mL (hold-up volume)
technique(s)
sterile filtration: suitable
H
164 mm
diam.
12.5 cm
filter capacity
250 mL
filter diam.
73 mm
filtration area
40 cm2
matrix
Express PLUS™
pore size
0.22 μm pore size
fitting
inlet funnel, outlet flask (Receiver)
shipped in
ambient
storage temp.
no temp limit
General description
Las unidades de filtración estériles Stericup® y Steritop® utilizan membranas de alto rendimiento fiables para una esterilización por filtración óptima de los reactivos, las disoluciones tampón y los medios. Estos dispositivos combinan caudales y rendimiento superiores con una baja adsorción inespecífica y un diseño estable sin puntas.
Esta unidad de filtración Stericup® con membrana de polietersulfona (PES) y tamaño de poro de 0,22 μm combina una tapa de botella de 250 ml con un matraz receptor de 250 ml y una tapa para procesamiento y almacenamiento.
Application
- Medios de cultivo de tejidos +/- aditivos
- Amortiguadores
- Disoluciones biológicas
Features and Benefits
- El embudo de filtración Quick Release se desconecta de la botella receptora con solo un cuarto de vuelta—reduciendo así la probabilidad de derrame y haciendo que sea más fácil manipularlo sobre la mesa de trabajo o debajo de la campana.
- La identificación del producto en negrita en el embudo de filtración Steritop® muestra claramente las características de la membrana, como su composición y el tamaño de los poros, una selección de productos rápida y apropiada para la aplicación.
- El tapón de cierre mediante un clic está bridado y modificado para permitir un fuerte agarre en condiciones secas—y un tope táctil confirma el cierre seguro. Preocúpese menos de la posibilidad de derramamiento o contaminación de los contenidos esterilizados.
- Una amplia superficie esmerilada de escritura en la botella receptora y el color más claro del tapón facilitan el etiquetado claro para mejorar la legibilidad y reducir los frotis ilegibles.
- La tapa descansa sobre el lateral para evitar el riesgo de contaminación.
- Las botellas receptoras son apilables para ahorrar espacio de almacenamiento.
Legal Information
Still not finding the right product?
Explore all of our products under Sistema de filtración al vacío Millipore® Stericup® Quick Release
Elija entre una de las versiones más recientes:
¿Ya tiene este producto?
Encuentre la documentación para los productos que ha comprado recientemente en la Biblioteca de documentos.
-
Is the Stericup® Quick Release Vacuum Filtration System certified RNase/DNase-Free?
1 answer-
Helpful?
-
-
Buongiorno. I filtri resistono ad inchiostri UV a base acrilica, all'acetone e all'alcool isopropilico?
1 answer-
This product is composed of polyethersulfone and polystyrene materials. Hence, it is not recommended for use with acetone.
Please see the chemical compatibility chart: https://www.sigmaaldrich.com/technical-documents/technical-article/analytical-chemistry/filtration/chemical-compatibility-of-filter-components?utm_source=redirect&utm_medium=promotional&utm_campaign=FilterChemicalCompatibilityHowever, the product's compatibility with acrylic-based UV inks and isopropyl alcohol has not been validated.
Please see the user guide for more information:
https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/337/924/20199204w-ms.pdfHelpful?
-
-
What tubing can be used with Stericup® and Steritop® vacuum filters?
What tubing can be used with Stericup® and Steritop® vacuum filters?
1 answer-
Stericup® and Steritop® filters are fitted with a 1/4” tube adapter. We recommend 3/16 in. (4.8 mm) ID silicone tubing to fit this tube adapter: https://www.sigmaaldrich.com/product/mm/xx7100004
Helpful?
-
-
Are Stericup® and Steritop® filters GMP-certified?
1 answer-
No, Stericup® and Steritop® devices are not GMP-approved.
Helpful?
-
-
What is the maximum processing volume for Stericup® and Steritop® filters? Can I add more volume after initial filtration?
1 answer-
All Stericup® and Steritop® filters were developed for single-use application. The maximum volume is dependent on the receiver bottle capacity. Stericup® filters come with a receiver bottle with a specified maximum volume. For Steritop® filters, the maximum recommended volume is 2 L with an appropriate-sized receiver bottle. Larger processing volumes can result in membrane fouling, increase chance of spillage and contamination, and lead to possible airlock*.
*Note that devices with PES membranes can airlock if filtered to dryness. Some liquid must be left on top of the filter, then the system stopped, receiver changed or emptied, and more sample volume added before restart. This increases the chance of spillage.Helpful?
-
-
What Bacterial Endotoxins Test is used?
1 answer-
The test method used complies with USP 85 and EP 2.6.14 to ensure the absence of endotoxin in our products.
Helpful?
-
-
What leachables are detected during filtration processes using Stericup® filters?
1 answer-
We have not performed extractables/leachables studies for our Stericup® filters.
Helpful?
-
-
What is the maximum vacuum pressure for filtration?
1 answer-
700 mm Hg (27.5” Hg, 0.9 bar) differential vacuum at 25 °C (77 °F)
Helpful?
-
-
Can we filter oil-type liquids with Stericup® filters?
Can we filter oil-type liquids with Stericup® filters?
1 answer-
Likely not, due to pore size/possible fouling/chemical compatibility. Refer to our guide, "Chemical compatibility of filter components": https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/302/855/membrane-chemical-compatibility-ds2532en-ms.pdf
Helpful?
-
-
What is the expiration date of Stericup® and Steritop® filters?
1 answer-
Typically 3 years from sterilization.
Helpful?
-