While the user interface (screen) of the RQflex 20 is different from that of the RQflex 10 and 10 plus, most of the test strips that had been provided for the 10/10 plus can be used on the RQflex 20. In fact, all test strips that had fit RQflex 10 and 10 plus, as long as they are not expired, are compatible with RQflex 20, except the following:
- Product 1.16993, calcium Test (2.5-45.0 mg/L Ca++)
- Product 1.17942, phosphate Test (0.1-5.0 mg/L PO4 3-)
- Product 1.17945, potassium Test (1.0-25.0 mg/L K+)
Zaloguj się, aby wyświetlić ceny organizacyjne i kontraktowe.
Informacje o tej pozycji
UNSPSC Code:
41116116
NACRES:
NA.24
Nazwa produktu
Test wapnia dla RQflex® 10 i RQflex® 10 plus, reflectometric, 2.5-45.0 mg/L (Ca), for use with REFLECTOQUANT®
description
quantitative
Quality Segment
specific analyte(s)
calcium
measuring range
2.5-45.0 mg/L (Ca)
compatibility
RQflex® 10 Plus, RQflex® 10, RQflex® 20, for use with REFLECTOQUANT®
detection method
reflectometric
storage temp.
15-25°C
General description
Metoda: reflektometryczna z paskami testowymi i odczynnikami 2,5 - 45,0 mg/l Ca Reflectoquant®
The Reflectoquant Calcium test includes test strips and all necessary additional reagents and is suitable for rapid determination of calcium in aqueous media. Calcium is an important parameter for boiling and cooling water. If the water contains calcium, deposits can be found in the plants and this can influence the function of the machines. Calcium also needs to be analyzed in drinking water and especially in beer. In the presence of hydrogen peroxide calcium ions react with glyoxal-bis(2-hydroxyanil) to form a red complex that is determined reflectometrically on the RQflex . For pre-testing a sample in an economic and quick way to estimate the calcium concentration and a possible dilution factor prior to using this quantitative method we recommend our semi-quantitative MQuant test strips for calcium. Further quantitative determination methods can be found with our Spectroquant tests.
Analysis Note
function control: conforms
Assay
- standard 1:
4.4 - 5.6 mg/l
- standard 2: 13.5 - 16.5 mg/l
- standard 3: 25.0 - 31.0 mg/l
Assay
- standard 1:
4.4 - 5.6 mg/l
- standard 2: 13.5 - 16.5 mg/l
- standard 3: 25.0 - 31.0 mg/l
Legal Information
REFLECTOQUANT is a registered trademark of Merck KGaA, Darmstadt, Germany
RQFLEX is a registered trademark of Merck KGaA, Darmstadt, Germany
Ta strona może zawierać tekst przetłumaczony maszynowo.
signalword
Danger
Hazard Classifications
Eye Dam. 1 - Met. Corr. 1 - Ox. Liq. 3 - Skin Corr. 1A
Klasa składowania
5.1B - Oxidizing hazardous materials
hcodes
Wybierz jedną z najnowszych wersji:
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
-
Bonjour, J'ai le RQ-Felx 20, tous les action réalisé par le 10 et 10plus sont il faisable par le 20 ? une bandelette dédier au 10 et 10 plus est-elle utilisable pour le 20 ? Merci et bonne journée
1 answer-
Helpful?
-