Durapore® 0.45 µm, Opticap® XL Capsule

Opticap® XL 10, cartridge nominal length 13.2 in. (33.5 cm), filter diam. 4.2 in. (10.7 cm)

Synonim(y):

Durapore® 0.45 µm, Opticap® XL Capsule

Zaloguj się, aby wyświetlić ceny organizacyjne i kontraktowe.

Informacje o tej pozycji

UNSPSC Code:
23151806


Nazwa produktu

Opticap® XL 10 CapsuleMedia: Sterile Durapore® 0.45 µm, Chemistry: Hydrophilic Polyvinylidene Fluoride (PVDF)Inlet: 3/4 in. Sanitary FlangeOutlet: 3/4 in. Sanitary FlangeMax Operating Pressure: 80 psi @ 25 °C

material

polyester support, polyethylene support, polypropylene filter, polypropylene housing, silicone outer O-ring, silicone-coated EPDM rubber lead O-ring

reg. compliance

FDA 21CFR177-182 (all component materials)

sterility

sterile; irradiated

product line

Opticap® XL 10

feature

hydrophilic

manufacturer/tradename

Opticap®

parameter

≤15.0 mL/min air diffusion at 3.9 bar (56 psig) and 23 °C (in water), 1 bar max. inlet pressure (15 psi) at 80 °C, 2.1 bar max. differential pressure (30 psid) at 4-40 °C (Reverse:), 2.8 bar max. inlet pressure (40 psi) at 60 °C, 25 °C max. inlet temp., 4.1 bar max. differential pressure (60 psid) at 25 °C (Forward:), 5.5 bar max. inlet pressure (80 psi) at 23 °C, 80 psig max. inlet pressure

technique(s)

bioburden reduction: suitable

cartridge nominal length

13.2 in. (33.5 cm)

device L

33.5 cm (13.2 in.)

device size

10 in.

filter diam.

4.2 in. (10.7 cm)

filter filtration area

0.62 m2

inlet connection diam.

3/4 in.

inlet to outlet width

33.5 cm (13.2 in.)

outlet connection diam.

3/4 in.

impurities

≤0.25 EU/mL bacterial endotoxins (LAL test, aqueous extraction), <0.25 EU/mL bacterial endotoxins (LAL test, sample aqueous extraction)

matrix

Durapore®

pore size

0.45 μm pore size

bubble point

≥1930 mbar (28 psig), air with water at 23 °C

fitting

1/4 in. fitting (vent/drain hose barb with double O-ring seal), (19 mm (3/4 in.) Sanitary Flange Inlet and Outlet)

General description

Device Configuration: Capsule

Preparation Note

Sterilization Method:
Gamma compatible to 40 kGy. 3 autoclave cycles of 60min at 126°C; not in-line steam sterilizable

This product was manufactured with a Durapore® membrane which meets the criteria for a "non-fiber releasing" filter as defined in 21CFR210.3(b)(6).

Analysis Note

Gravimetric Extractables: ≤ 25 mg per capsule after 24 hours in water at controlled room temperature.
Will meet the USP Oxidizable Substances Test requirements after a water flush of ≥ 1500 mL

Other Notes

Directions for Use:

  • Organism Retention: Microorganism
  • Mode of Action: Filtration (size exclusion)
  • Application: BioProcessing
  • Intended Use: Reduction or removal of microorganism/bioburden
  • Instructions for Use: Please refer installation section of Opticap® XL Capsules and Opticap® XLT Capsules user guide
  • Storage Statement: Store in dry location
  • Disposal Statement: Dispose of in accordance with applicable federal, state and local regulations

Legal Information

Durapore is a registered trademark of Merck KGaA, Darmstadt, Germany
OPTICAP is a registered trademark of Merck KGaA, Darmstadt, Germany
Ta strona może zawierać tekst przetłumaczony maszynowo.


Still not finding the right product?




Zgodność z GMP dzięki programowi Emprove®

Przyspiesz kwalifikacje, zachowaj zgodność i oszczędzaj czas dzięki Emprove® Dossiers.

Pobierz Dossiers

Uwaga: zostaniesz przeniesiony do EmproveSuite.com. Twoja nazwa użytkownika i hasło są niezależne od SigmaAldrich.com i mogą się różnić.

Zarejestruj się za darmo i uzyskaj dostęp do Emprove® Dossiers w ciągu kilku minut. Uaktualnij swój plan subskrypcji, aby odblokować zaawansowaną i spersonalizowaną zawartość w Emprove® Suite.

Nie znalazłeś tego, czego szukasz? Odwiedź stronę pomocy programu Emprove®..

Masz już ten produkt?

Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.

Odwiedź Bibliotekę dokumentów

Masz trudności ze znalezieniem produktu lub numeru serii/partii?

Podczas wyświetlania na stronie internetowej numery produktów są łączone z wielkością przesyłki / ilością (przykład: T1503-25G). Pamiętaj że w polu „Numer produktu” należy podać TYLKO numer produktu (przykład: T1503).

Przykład:

T1503
Numer produktu
-
25G
Wielkość opakowania/ilość

Przykłady dodatkowe:

705578-5MG-PW

PL860-CGA/SHF-1EA

MMYOMAG-74K-13

1000309185

wpisz jako 1.000309185)

Mieć kłopoty? Poczuj się swobodnie w kontakcie Pomoc techniczna dla wsparcia



Questions

Reviews

No rating value

Active Filters