Conditions | |
|---|---|
| column | SP-2560, 100 m x 0.25 mm I.D., 0.20 µm (24056) |
| oven | 140 °C (5 min), 8 °C/min to 180 °C, 4 °C/min to 210 °C, 20 °C/min to 250 °C (7 min) |
| carrier gas | hydrogen, 40 cm/sec |
| sample | Supelco 37-Component FAME Mix (47885-U), analytes at concentrations indicated in methylene chloride |
| injection | 1 µL, 100:1 split |
| liner | 4 mm I.D., cup design |
| inj. temp. | 250 °C |
| detector | FID, 250 °C |
Description | |
|---|---|
| Categories | Analytical Chromatography, GC Applications, Fatty Acid Methyl Esters (FAMEs) |
| Featured Industry | Agriculture Clinical Food and Beverages |
| Legal Information | SP is a trademark of Sigma-Aldrich Co. LLC |
| Other Notes | Complementary products: Syringes Vials Gas Purification and Gas Management Inlet Liners, Glass Wool, and Specialized Hand Tools Septa and Specialized Hand Tools Column Ferrules, Nuts, and Specialized Hand Tools |
| suitability | application for GC |