The low value observed in the sample may be due to several factors. Since the standard curve is working well, it is likely that the lysis buffer and/or homogenization condition used is not optimized for assay reading. To verify the possible presence of inhibiting substances, one could add a known amount of ATP from a standard solution to the homogenate to see if the reading increases accordingly. It's possible that the lysis buffer used was not optimal to yield an adequate amount of ATP for detection. Currently, there is no recommended cardiac tissue preparation protocol for use with this kit. As the kit measures only soluble ATP, it is highly recommended to clarify the homogenized extract by centrifugation to obtain a supernatant solution for use with this assay.
Przejdź do
Wybierz wielkość
| Rozmiar/SKU | Dostępność | Cena netto |
|---|---|---|
1 vial | Skontaktuj się z Obsługą Klienta, aby uzyskać informacje na temat dostępności | 405,00 zł |
5 vials | Skontaktuj się z Obsługą Klienta, aby uzyskać informacje na temat dostępności | 1160,00 zł |
10 vials | Skontaktuj się z Obsługą Klienta, aby uzyskać informacje na temat dostępności | 2170,00 zł 1844,50 zł |
Informacje o tej pozycji
405,00 zł
form
lyophilized powder
Quality Segment
technique(s)
activity assay: suitable, protein detection: suitable
storage temp.
−20°C
General description
Application
Preparation Note
1 of 1
Ta pozycja | |||
|---|---|---|---|
| description lyophilized powder | description lyophilized powder | description for ATP quantitation | description vial of ~1 mg ATP |
| technique(s) activity assay: suitable | technique(s) PCR: suitable | technique(s) - | technique(s) - |
| Quality Level 200 | Quality Level 200 | Quality Level 200 | Quality Level 200 |
| form lyophilized powder | form lyophilized powder | form - | form lyophilized powder |
| storage temp. −20°C | storage temp. −20°C | storage temp. −20°C | storage temp. −20°C |
Słowo ostrzegawcze
Danger
Kody zagrożeń
Ostrzeżenia
Hazard Classifications
Eye Dam. 1
Klasa składowania
11 - Combustible Solids
Co się dzieje?
WGK 3
Temperatura zapłonu (°F)
Not applicable
Temperatura zapłonu (°C)
Not applicable
PPE (środki ochrony indywidualnej)
Eyeshields, Gloves, type N95 (US)
Wybierz jedną z najnowszych wersji:
Masz już ten produkt?
Dokumenty związane z niedawno zakupionymi produktami zostały zamieszczone w Bibliotece dokumentów.
-
It seems that you are currently trying to perform the ATP assay (Sigma Cat# FLAA) on cardiac tissue. The kit appears to be working, as your standard curve is functioning, but you are obtaining very low numbers for your samples, significantly below your standard curve. Given that the heart should contain an abundance of ATP, you are concerned that your extraction protocol might be the issue. You are wondering if there is an extraction protocol available for this tissue for the mentioned kit.
1 answer-
Helpful?
-