The speed needs to be evaluated. However, we know that the maximum speed should never be exceeded. It is best to run below the maximum speed to not risk breakage in case the centrifuge has not been calibrated recently. Here are our specifications for the Amicon Ultra 0.2: Maximum relative centrifugal force Swinging bucket rotor 4,000 × g for concentration spin, 1,000 × g for recovery spin Fixed angle rotor 7,500 × g for concentration spin, 1,000 × g for recovery spin
NOTE: When spinning viscous solutions such as undiluted serum or plasma, do not exceed 5,400 x g. Furthermore, if you are trying to concentrate proteins, often, proteins do not like to be spun at a high speed or they can precipitate. Therefore, let's say your centrifuge is run at 2000 x g. This should work fine. It may take longer or it may not. This depends on initial and final desired volume when stopped.
Select a Size
| Size/SKU | Availability | Price |
|---|---|---|
24 pkg | Estimated to ship onDecember 08, 2026 | $217.00 |
About This Item
Quality Segment
material
polypropylene tube (for collection), regenerated cellulose membrane, styrene-butadiene housing
sterility
non-sterile
product line
Amicon®
packaging
pkg of 24 pkg
manufacturer/tradename
Amicon® Ultra
sustainability
Sustainability features
parameter
2 mL sample volume
technique(s)
Diafiltration: suitable, protein extraction: suitable, protein purification: suitable
L
119.7 mm
diam.
15.9 mm
filtration area
1 cm2
volume
2 mL
working volume
≤2 mL , 2 mL
pore size
10 kDa MWCO
shipped in
ambient
General description
Application
- to reduce large volume biofluids prior to extracellular vesicle (EV) isolation or to concentrate EVs
- Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids (DNA/RNA samples, either single- or double-stranded), microorganisms, column eluates, and purified samples
- Purification of macromolecular components found in tissue culture extracts and cell lysates, removal of primer, linkers, or molecular labels from a reaction mix, and protein removal before high-performance liquid chromatography (HPLC)
- Desalting, buffer exchange, or diafiltration
- Ultrafiltration with 10,000 Dalton molecular weight cutoff Ultracel® membrane.
- Concentration of biological samples containing antigens, antibodies, enzymes, nucleic acids, or microorganisms.
- Purification of macromolecular components found in tissue culture extracts or cell lysates.
- Primer removal, PCR cleanup, removal of linkers or macromolecular labels from a reaction mix, and protein removal prior to HPLC.
- Desalting, buffer exchange and protein dialysis.
- Size exclusion and protein fractionation.
Features and Benefits
- Amicon® Ultra 2ml 10K 24PK
- Concentrates 2 mL down to 15-70 μl
- 98% sample recovery
- Fast processing time - 10 to 30 minute spin times
- Engineered dead stop
- The reverse spin capability provides consistent recoveries
- Ultracel® regenerated cellulose membrane with 10 kDa NMWCO delivers retentate recovery of greater than 90%
- Concentrates 2 mL down to 15-70 μL
- 98% sample recovery
- Fast processing time: 10 to 30-minute spin times
- Engineered dead stop
- Reverse spin capability provides consistent recoveries
Other Notes
Legal Information
1 of 1
This Item | |||
|---|---|---|---|
| description sample volume 2 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 4 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 15 mL, regenerated cellulose membrane, MWCO 10 kDa | description sample volume 0.5 mL, regenerated cellulose membrane, MWCO 10 kDa |
| pore size 10 kDa MWCO | pore size 10 kDa MWCO | pore size 10 kDa MWCO | pore size 10 kDa MWCO |
| sterility non-sterile | sterility non-sterile | sterility non-sterile | sterility non-sterile |
| manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra | manufacturer/tradename Amicon® Ultra |
| parameter 2 mL sample volume | parameter 4 mL sample volume | parameter 15 mL sample volume | parameter 0.5 mL sample volume |
| material styrene-butadiene housing, polypropylene tube (for collection), regenerated cellulose membrane | material polypropylene tube, polyethylene cap, regenerated cellulose membrane, regenerated cellulose support, styrene-butadiene housing, polyethylene liner | material polypropylene tube, styrene-butadiene housing, regenerated cellulose support, polyethylene cap, polyethylene liner, regenerated cellulose membrane | material styrene-butadiene housing, regenerated cellulose membrane, polypropylene tube, polypropylene cap |
| working volume ≤2 mL | working volume ≤4 mL | working volume ≤15 mL | working volume 0.5 mL , ≤500 μL |
Still not finding the right product?
Explore all of our products under Amicon® Ultra Centrifugal Filter, 10 kDa MWCO
Choose from one of the most recent versions:
Already Own This Product?
Find documentation for the products that you have recently purchased in the Document Library.
-
At what speed should the Amicon® Ultra Centrifugal Filter, 10 kDa MWCO UFC201024 be used? Besides, what is the maximun g-force supported by the Amicon?
1 answer-
Helpful?
-
-
Can Amicon® Ultra Centrifugal Filter, 10 kDa MWCO UFC201024 to be in the Centrifuge as 4 mL Centrifugal Filter without adaptors?We have adaptors for 15 mL TUBES
1 answer-
Amicon Ultra-2 devices should be used in a centrifuge with swinging bucket or fixed angle rotor with wells/carriers that can accommodate 17 mm × 100 mm tubes. This is the same well/carrier size that is recommended for the Amicon® Ultra-4 devices. To avoid damage to the device during centrifugation, make sure it is properly assembled and seated at the bottom of the rotor. The rim of the concentrate collection tube should be inside the rotor well. Check clearance before spinning.
Helpful?
-
-
Hello. Does these 2 ml 10 kDa filters fit into the same centrifuge rotor as the 0.5 ml 10 kDa amicon filters?
1 answer-
No, the Amicon-0.5 devices require a microfuge, while the Amicon-2 devices require a centrifuge with adapters for 5 ml tubes.
Helpful?
-