The reverse spin process requires inversion of the ultrafiltration device into an appropriate collection tube in order to maximize sample recovery. To recover the concentrated solute, place the Amicon® Ultra filter device upside down in a clean microcentrifuge tube. Place in centrifuge, aligning open cap towards the center of the rotor; counterbalance with a similar device. Spin for 2 minutes at 1,000 × g to transfer the concentrated sample from the device to the tube. The ultrafiltrate can be stored in the centrifuge tube. Please see the link below to view and instructional video:
https://www.youtube.com/watch?v=jXnKT6xHY-8
Please see the link below to review additional details and instructions in the product User Guide:
https://www.sigmaaldrich.com/deepweb/assets/sigmaaldrich/product/documents/105/563/pr05485w-ug-ms.pdf
The following link includes additional helpful information regarding the selection and use of Amicon UltraFiltration Devices:
https://www.sigmaaldrich.com/products/protein-biology/protein-sample-prep/amicon-centrifugals#ultracentrifugal