GC Analysis of Resveratrol in Red Wine on SLB®-5ms after SPME using 85 μm Polyacrylate Fiber

Materiali
Accessorio
N° Catalogo
Descrizione
Determinazione del prezzo
Vials, screw top with phenolic open-top cap, pre-assembled
volume 4 mL, clear glass vial, O.D. × H 15 mm × 45 mm, tan PTFE/silicone septumProdotti correlati
used together
N° Catalogo
Descrizione
Determinazione del prezzo
Gruppo fibre SPME poliacrilato (PA)
85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, white hubGruppo fibre SPME poliacrilato (PA)
85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Autosampler, pk of 3, white hubGruppo fibre SPME poliacrilato (PA)
85 μm Polyacrylate, Fused Silica (1 cm), needle size 23 ga, Autosampler, pk of 3, white hubCONDITIONS
sample/matrix
3 mL of red wine (California merlot) diluted 3:1 in 12% ethanol:water
SPME fiber
85 μm polyacrylate FS 23 gauge auto (57294-U)
SPME fiber
85 μm polyacrylate FS 24gauge auto (57305)
SPME fiber
85 μm polyacrylate FS 24gauge manual (57304)
extraction
immersion for 15 min at room temperature while stirring at 400 rpm, fiber gently blotted with a Kimwipe to remove excess water, (then derivatization by immersion in a 4 mL vial containing 5 μL of Sylon-BFT (fresh vial for each analysis than was allowed to equilabrate 60-90 minutes prior to use) for 20 min)
desorption process
280 °C for 2 min
column
SLB-5ms; 30 m x 0.25 mm I.D., 0.25 μm (28471-U)
oven
100 °C (1 min), 10 °C/min to 325 °C (3 min)
MSD interface
325 °C
scan range
m/z 40-450
carrier gas
helium, 1 mL/min, constant
liner
0.75 mm I.D., SPME
Descrizione
Note legali
SLB is a registered trademark of Merck KGaA, Darmstadt, Germany